A-8
DINAMAP
®
ProCare Monitor Service Manual
Revision C
2009381-001
Technical Specifications:
Specifications
‡The Masimo SET
®
SpO
2
parameter has been validated for low perfusion accuracy in bench top testing
against a Biotek Index 2 stimulator and Masimo’s simulator with signal strengths of greater than 0.02% and
a % transmission of greater than 5% for saturations ranging from 70 to 100%. This variation equals plus or
minus, one standard deviation.Plus or minus one standard deviation encompasses 68% of the population.
Masimo
®
Sensor Accuracy
Sensor Model
SpO2 Range
70% - 100%
LNOP
LNOP-ADT
± 2 digits
LNOP-ADT Long
± 2 digits
LNOP-PDT
± 2 digits
LNOP-NEO
± 3 digits
LNOP-NEO PT
± 3 digits
LNOP-DCI (reusable)
± 2 digits
LNOP-DCSC (reusable)
± 2 digits
LNOP-DCIP (reusable)
± 2 digits
NRI25 (reusable)
± 2 digits
LNOP-DC195 (reusable)
± 2 digits
Resolution
Saturation (% SpO
2
)
1%
Pulse Rate (bpm)
1
Low Perfusion Performance
>0.02% Pulse Amplitude
Saturation (% SpO
2
) ±2 digits
and % Transmission >5%
Pulse Rate ±3 digits
Interfering Substances
Carboxyhemoglobin may erroneously increase readings. The level of increase is
approximately equal to the amount of carboxyhemoglobin present. Dyes, or any
substance containing dyes, that change usual arterial pigmentation may cause
erroneous readings.
Sensor Light Source
Wavelength
Infrared: 905 nm (nominal)
Red: 660 nm (nominal)
Power Dissipation
Infrared: 22.5 mW (max)
Red: 27.5 mW (max)
Factory Default Settings
SpO
2
(%)
HIGH: 100
SpO
2
(%)
LOW: 90
Sensitivity Mode
2 (for low perfusion-Default)
Averaging Time
12 seconds
FastSAT Mode
0 (for Off)
Masimo Patents
5,482,036; 5,490,505; 5,632,272; 5,685,299; 5,758,644; 5,769,785; 6,002,952;
6,036,642; 6,067,462; 6,206,830; 6,157,850, and international equivalents.
Содержание DINAMAP ProCare
Страница 2: ......
Страница 11: ...Revision C DINAMAP ProCare Monitor Service Manual 1 1 2009381 001 1 Introduction ...
Страница 12: ...1 2 DINAMAP ProCare Monitor Service Manual Revision C 2009381 001 For your notes ...
Страница 23: ...Revision C DINAMAP ProCare Monitor Service Manual 2 1 2009381 001 2 Equipment Overview ...
Страница 24: ...2 2 DINAMAP ProCare Monitor Service Manual Revision C 2009381 001 For your notes ...
Страница 38: ...2 16 DINAMAP ProCare Monitor Service Manual Revision C 2009381 001 Equipment Overview Functional Description ...
Страница 39: ...Revision C DINAMAP ProCare Patient Monitor Service Manual 3 1 2009381 001 3 Installation ...
Страница 40: ...3 2 DINAMAP ProCare Patient Monitor Service Manual Revision C 2009381 001 For your notes ...
Страница 53: ...Revision C DINAMAP ProCare Monitor Service Manual 4 1 2009381 001 4 Maintenance ...
Страница 54: ...4 2 DINAMAP ProCare Monitor Service Manual Revision C 2009381 001 For your notes ...
Страница 71: ...Revision C DINAMAP ProCare Monitor Service Manual 5 1 2009381 001 5 Troubleshooting ...
Страница 72: ...5 2 DINAMAP ProCare Monitor Service Manual Revision C 2009381 001 For your notes ...
Страница 76: ...5 6 DINAMAP ProCare Monitor Service Manual Revision C 2009381 001 Troubleshooting Alarm Code Interpretation ...
Страница 77: ...Revision C DINAMAP ProCare Monitor Service Manual 6 1 2009381 001 6 Parts List Drawings and Replacement ...
Страница 78: ...6 2 DINAMAP ProCare Monitor Service Manual Revision C 2009381 001 For your notes ...
Страница 99: ...Revsion C DINAMAP ProCare Monitor Service Manual A 1 2009381 001 A Technical Specifications ...
Страница 100: ...A 2 DINAMAP ProCare Monitor Service Manual Revision C 2009381 001 For your notes ...
Страница 108: ...A 10 DINAMAP ProCare Monitor Service Manual Revision C 2009381 001 Technical Specifications Specifications ...
Страница 109: ...Revision C DINAMAP ProCare Monitor Service Manual B 1 2009381 001 B Appropriate Use of NIBP Simulators ...
Страница 110: ...B 2 DINAMAP ProCare Monitor Service Manual Revision C 2009381 001 Appropriate Use of NIBP Simulators For your notes ...
Страница 115: ...Revision C DINAMAP ProCare Monitor Service Manual C 1 2009381 001 C Electromagnetic Compatibility EMC ...
Страница 116: ...C 2 DINAMAP ProCare Monitor Service Manual Revision C 2009381 001 Electromagnetic Compatibility EMC For your notes ...
Страница 133: ......