Revision C
DINAMAP
®
ProCare Monitor Service Manual
3-7
2009381-001
Installation:
Configuring Your ProCare Monitor
Pressing the
Menu
button cycles through all the configuration option pages.
After all options pages have been displayed, the display returns to the 1
st
configuration mode page (displaying
CFG
). You can use the
+
and
-
buttons to
make changes to settings. After making changes, simply cycle the power to
return to normal operation (clinical) mode. Changes are automatically retained.
The Menu selections appear in the following order. Refer the each manual
section for settings options.
NOTE:
Menu selections for a Masimo unit have three SpO
2
settings. Refer to the
Masimo SpO
2
section for options.
Setting:
Window
LED Display
• Inflate
Pressure:
Diastolic
XXX (numeric)
• SpO
2
Mode:
SpO
2
• SpO
2
Sat:
SpO
2
• SpO
2
* Sensitivity:
SpO
2
• Temp:
°
C or
°
F
• Year:
Systolic
• Month:
MAP/Cuff
• Day:
Diastolic
• Hour:
min
• Minute:
min
• Mode:
Systolic
*Masimo units only
Setting the Date and Time –
To set the date and time on the ProCare
Monitor,
you must access the configuration mode. Press
Menu
to skip the default settings
that do not require changes. The following list shows the windows in which the
date/time settings appear.
Setting:
Window
LED Display
• Year:
Systolic
• Month:
MAP/Cuff
• Day:
Diastolic
• Hour:
min
• Minute:
min
Procedures
1. Press the
Menu
button to move from one setting to another. Use the
+/-
buttons to increment or decrement the setting.
NOTE
:
For the date and time to be saved, you must advance the menu
through the minute setting.
2. To exit the configuration mode, press the
On/Off
button.
3. To continue with other changes, press the
Menu
button.
CFG
will appear in
the Systolic window. To change parameter settings, press the
Menu
button
and select the parameter function. To change alarm settings, press the
Alarms
button.
Содержание DINAMAP ProCare
Страница 2: ......
Страница 11: ...Revision C DINAMAP ProCare Monitor Service Manual 1 1 2009381 001 1 Introduction ...
Страница 12: ...1 2 DINAMAP ProCare Monitor Service Manual Revision C 2009381 001 For your notes ...
Страница 23: ...Revision C DINAMAP ProCare Monitor Service Manual 2 1 2009381 001 2 Equipment Overview ...
Страница 24: ...2 2 DINAMAP ProCare Monitor Service Manual Revision C 2009381 001 For your notes ...
Страница 38: ...2 16 DINAMAP ProCare Monitor Service Manual Revision C 2009381 001 Equipment Overview Functional Description ...
Страница 39: ...Revision C DINAMAP ProCare Patient Monitor Service Manual 3 1 2009381 001 3 Installation ...
Страница 40: ...3 2 DINAMAP ProCare Patient Monitor Service Manual Revision C 2009381 001 For your notes ...
Страница 53: ...Revision C DINAMAP ProCare Monitor Service Manual 4 1 2009381 001 4 Maintenance ...
Страница 54: ...4 2 DINAMAP ProCare Monitor Service Manual Revision C 2009381 001 For your notes ...
Страница 71: ...Revision C DINAMAP ProCare Monitor Service Manual 5 1 2009381 001 5 Troubleshooting ...
Страница 72: ...5 2 DINAMAP ProCare Monitor Service Manual Revision C 2009381 001 For your notes ...
Страница 76: ...5 6 DINAMAP ProCare Monitor Service Manual Revision C 2009381 001 Troubleshooting Alarm Code Interpretation ...
Страница 77: ...Revision C DINAMAP ProCare Monitor Service Manual 6 1 2009381 001 6 Parts List Drawings and Replacement ...
Страница 78: ...6 2 DINAMAP ProCare Monitor Service Manual Revision C 2009381 001 For your notes ...
Страница 99: ...Revsion C DINAMAP ProCare Monitor Service Manual A 1 2009381 001 A Technical Specifications ...
Страница 100: ...A 2 DINAMAP ProCare Monitor Service Manual Revision C 2009381 001 For your notes ...
Страница 108: ...A 10 DINAMAP ProCare Monitor Service Manual Revision C 2009381 001 Technical Specifications Specifications ...
Страница 109: ...Revision C DINAMAP ProCare Monitor Service Manual B 1 2009381 001 B Appropriate Use of NIBP Simulators ...
Страница 110: ...B 2 DINAMAP ProCare Monitor Service Manual Revision C 2009381 001 Appropriate Use of NIBP Simulators For your notes ...
Страница 115: ...Revision C DINAMAP ProCare Monitor Service Manual C 1 2009381 001 C Electromagnetic Compatibility EMC ...
Страница 116: ...C 2 DINAMAP ProCare Monitor Service Manual Revision C 2009381 001 Electromagnetic Compatibility EMC For your notes ...
Страница 133: ......