Configuring Your ProCare Monitor
Inflation Pressure Default Setting –
Procedures
1. Enter the configuration mode: with the Monitor off, press and hold the
Menu
button at the same time as pressing and holding the
On/Off
button for 3
seconds, or press
Menu
until the Inflate Pressure is lit on the display and the
pressure is displayed in the Diastolic window.
2. Use the
+/-
buttons to increment or decrement the inflate pressure default
setting. Increments are 5mmhg from 100mmhg to 250mmhg.
3. To exit the configuration mode, turn the unit off. To continue with additional
configuration settings, press
Menu
.
Alarm Default Settings –
Procedures
1. Enter the configuration mode: with the Monitor off, press and hold the
Menu
button at the same time as pressing and holding the
On/Off
button for 3
seconds. After the unit enters the configuration mode, press
Alarms
. At any
point in the configuration mode menu, Alarms default can be selected.
2. To set or change the default setting, press the Alarms button to select alarm
setting. Use the
+/-
buttons to increment or decrement the individual
settings.
NOTE
:
For the Alarms default setting to be saved, you must advance the
menu through the SPO2 settings.
3. To exit the configuration mode, turn the unit off. To continue with additional
configuration settings, press
Menu
.
SpO
2
Configuration Settings –
Procedure for units with Nellcor® Technology
(Refer to the Nellcor® section for options)
1. Enter the configuration mode: with the Monitor off, press and hold the
Menu
button at the same time as pressing and holding the
On/Off
button for 3
seconds.
2. Press the menu button until
n
0d
appears in the
Pulse Rate
window.
3. Use the
+/-
buttons to select the option.
4. Press the
Menu
button once.
SAt
appears in the
Pulse Rate
window.
5. Use the
+/-
buttons to select the option.
6. To exit the configuration mode, turn the unit off. To continue with additional
configuration settings, press
Menu
.
Содержание DINAMAP ProCare
Страница 2: ......
Страница 11: ...Revision C DINAMAP ProCare Monitor Service Manual 1 1 2009381 001 1 Introduction ...
Страница 12: ...1 2 DINAMAP ProCare Monitor Service Manual Revision C 2009381 001 For your notes ...
Страница 23: ...Revision C DINAMAP ProCare Monitor Service Manual 2 1 2009381 001 2 Equipment Overview ...
Страница 24: ...2 2 DINAMAP ProCare Monitor Service Manual Revision C 2009381 001 For your notes ...
Страница 38: ...2 16 DINAMAP ProCare Monitor Service Manual Revision C 2009381 001 Equipment Overview Functional Description ...
Страница 39: ...Revision C DINAMAP ProCare Patient Monitor Service Manual 3 1 2009381 001 3 Installation ...
Страница 40: ...3 2 DINAMAP ProCare Patient Monitor Service Manual Revision C 2009381 001 For your notes ...
Страница 53: ...Revision C DINAMAP ProCare Monitor Service Manual 4 1 2009381 001 4 Maintenance ...
Страница 54: ...4 2 DINAMAP ProCare Monitor Service Manual Revision C 2009381 001 For your notes ...
Страница 71: ...Revision C DINAMAP ProCare Monitor Service Manual 5 1 2009381 001 5 Troubleshooting ...
Страница 72: ...5 2 DINAMAP ProCare Monitor Service Manual Revision C 2009381 001 For your notes ...
Страница 76: ...5 6 DINAMAP ProCare Monitor Service Manual Revision C 2009381 001 Troubleshooting Alarm Code Interpretation ...
Страница 77: ...Revision C DINAMAP ProCare Monitor Service Manual 6 1 2009381 001 6 Parts List Drawings and Replacement ...
Страница 78: ...6 2 DINAMAP ProCare Monitor Service Manual Revision C 2009381 001 For your notes ...
Страница 99: ...Revsion C DINAMAP ProCare Monitor Service Manual A 1 2009381 001 A Technical Specifications ...
Страница 100: ...A 2 DINAMAP ProCare Monitor Service Manual Revision C 2009381 001 For your notes ...
Страница 108: ...A 10 DINAMAP ProCare Monitor Service Manual Revision C 2009381 001 Technical Specifications Specifications ...
Страница 109: ...Revision C DINAMAP ProCare Monitor Service Manual B 1 2009381 001 B Appropriate Use of NIBP Simulators ...
Страница 110: ...B 2 DINAMAP ProCare Monitor Service Manual Revision C 2009381 001 Appropriate Use of NIBP Simulators For your notes ...
Страница 115: ...Revision C DINAMAP ProCare Monitor Service Manual C 1 2009381 001 C Electromagnetic Compatibility EMC ...
Страница 116: ...C 2 DINAMAP ProCare Monitor Service Manual Revision C 2009381 001 Electromagnetic Compatibility EMC For your notes ...
Страница 133: ......