4-12
DINAMAP
®
ProCare Monitor Service Manual
Revision C
2009381-001
Maintenance:
Parameter Level Functional Testing
Pressure Transducer Verification
NOTE:
To enter Service Mode press and hold the
Cycle
button while pressing
the
On/Off
button.
1. Turn unit on and enter Service Mode.
2. The
min
window should display
0
.
3. Open the valve on the inflation bulb and remove all pressure from the
system (manometer reads zero).
4. Press
Cycle
button and
1
should appear in the
min
window.
5. Use the inflation bulb to inflate the cuff, hose and pressure indicator setup to
200 mmHg.
6. Record and verify the pressure reading that appears in the top
Systolic
window.
7. Record and verify the pressure reading that appears in the top
Diastolic
window.
8. Use the valve on the bulb to reduce pressure to 150 mmHg.
9. Record and verify the pressure reading that appears in the top
Systolic
window.
10. Record and verify the pressure reading that appears in the top
Diastolic
window.
11. Use the valve on the bulb to reduce pressure to 100 mmHg.
12. Record and verify the pressure reading that appears in the top
Systolic
window.
13. Record and verify the pressure reading that appears in the top
Diastolic
window.
14. Use the valve on the bulb to reduce pressure to 50 mmHg.
15. Record and verify the pressure reading that appears in the top
Systolic
window.
16. Record and verify the pressure reading that appears in the top
Diastolic
window.
Pressure Transducer Calibration
Perform only if Pressure Transducer Verification is out of tolerance as
specified in Test Results Table at the end of this section.
NOTE:
To enter service mode press and hold
Cycle
button while pressing the
On/Off
button.
1. Turn the ProCare Monitor on and enter Service Mode.
2. The
min
window should display
0
.
3. Open valve on bulb to open pressure system to atmosphere.
4. Verify the manometer reads zero.
5. Press
Cycle
button and
1
should appear in the
min
window.
Содержание DINAMAP ProCare
Страница 2: ......
Страница 11: ...Revision C DINAMAP ProCare Monitor Service Manual 1 1 2009381 001 1 Introduction ...
Страница 12: ...1 2 DINAMAP ProCare Monitor Service Manual Revision C 2009381 001 For your notes ...
Страница 23: ...Revision C DINAMAP ProCare Monitor Service Manual 2 1 2009381 001 2 Equipment Overview ...
Страница 24: ...2 2 DINAMAP ProCare Monitor Service Manual Revision C 2009381 001 For your notes ...
Страница 38: ...2 16 DINAMAP ProCare Monitor Service Manual Revision C 2009381 001 Equipment Overview Functional Description ...
Страница 39: ...Revision C DINAMAP ProCare Patient Monitor Service Manual 3 1 2009381 001 3 Installation ...
Страница 40: ...3 2 DINAMAP ProCare Patient Monitor Service Manual Revision C 2009381 001 For your notes ...
Страница 53: ...Revision C DINAMAP ProCare Monitor Service Manual 4 1 2009381 001 4 Maintenance ...
Страница 54: ...4 2 DINAMAP ProCare Monitor Service Manual Revision C 2009381 001 For your notes ...
Страница 71: ...Revision C DINAMAP ProCare Monitor Service Manual 5 1 2009381 001 5 Troubleshooting ...
Страница 72: ...5 2 DINAMAP ProCare Monitor Service Manual Revision C 2009381 001 For your notes ...
Страница 76: ...5 6 DINAMAP ProCare Monitor Service Manual Revision C 2009381 001 Troubleshooting Alarm Code Interpretation ...
Страница 77: ...Revision C DINAMAP ProCare Monitor Service Manual 6 1 2009381 001 6 Parts List Drawings and Replacement ...
Страница 78: ...6 2 DINAMAP ProCare Monitor Service Manual Revision C 2009381 001 For your notes ...
Страница 99: ...Revsion C DINAMAP ProCare Monitor Service Manual A 1 2009381 001 A Technical Specifications ...
Страница 100: ...A 2 DINAMAP ProCare Monitor Service Manual Revision C 2009381 001 For your notes ...
Страница 108: ...A 10 DINAMAP ProCare Monitor Service Manual Revision C 2009381 001 Technical Specifications Specifications ...
Страница 109: ...Revision C DINAMAP ProCare Monitor Service Manual B 1 2009381 001 B Appropriate Use of NIBP Simulators ...
Страница 110: ...B 2 DINAMAP ProCare Monitor Service Manual Revision C 2009381 001 Appropriate Use of NIBP Simulators For your notes ...
Страница 115: ...Revision C DINAMAP ProCare Monitor Service Manual C 1 2009381 001 C Electromagnetic Compatibility EMC ...
Страница 116: ...C 2 DINAMAP ProCare Monitor Service Manual Revision C 2009381 001 Electromagnetic Compatibility EMC For your notes ...
Страница 133: ......