Revision C
DINAMAP
®
ProCare Monitor Service Manual
4-11
2009381-001
Maintenance:
Parameter Level Functional Testing
3. Consult the following diagram for pneumatic setup guidelines.
Leakage Testing
NOTE:
To enter Service Mode press and hold the
Cycle
button while pressing
the
On/Off
button.
1. Turn unit on and enter Service Mode.
2. Press
Cycle
button and
1
should appear in the
min
window.
3. Close the valve on the inflation bulb.
4. Using the inflation bulb, inflate the system to 210 mmHg.
5. Allow the system to stabilize for 5 seconds (it is normal to see some
decrease in pressure at this point).
6. Start the stopwatch and record the pressure value.
7. After 60 seconds record the pressure value.
8. The leakage rate is the difference between the first and second readings.
9. Record and verify the leakage rate.
10. Turn the ProCare Monitor off.
MANDREL
2 TUBE
BP CUFF
Содержание DINAMAP ProCare
Страница 2: ......
Страница 11: ...Revision C DINAMAP ProCare Monitor Service Manual 1 1 2009381 001 1 Introduction ...
Страница 12: ...1 2 DINAMAP ProCare Monitor Service Manual Revision C 2009381 001 For your notes ...
Страница 23: ...Revision C DINAMAP ProCare Monitor Service Manual 2 1 2009381 001 2 Equipment Overview ...
Страница 24: ...2 2 DINAMAP ProCare Monitor Service Manual Revision C 2009381 001 For your notes ...
Страница 38: ...2 16 DINAMAP ProCare Monitor Service Manual Revision C 2009381 001 Equipment Overview Functional Description ...
Страница 39: ...Revision C DINAMAP ProCare Patient Monitor Service Manual 3 1 2009381 001 3 Installation ...
Страница 40: ...3 2 DINAMAP ProCare Patient Monitor Service Manual Revision C 2009381 001 For your notes ...
Страница 53: ...Revision C DINAMAP ProCare Monitor Service Manual 4 1 2009381 001 4 Maintenance ...
Страница 54: ...4 2 DINAMAP ProCare Monitor Service Manual Revision C 2009381 001 For your notes ...
Страница 71: ...Revision C DINAMAP ProCare Monitor Service Manual 5 1 2009381 001 5 Troubleshooting ...
Страница 72: ...5 2 DINAMAP ProCare Monitor Service Manual Revision C 2009381 001 For your notes ...
Страница 76: ...5 6 DINAMAP ProCare Monitor Service Manual Revision C 2009381 001 Troubleshooting Alarm Code Interpretation ...
Страница 77: ...Revision C DINAMAP ProCare Monitor Service Manual 6 1 2009381 001 6 Parts List Drawings and Replacement ...
Страница 78: ...6 2 DINAMAP ProCare Monitor Service Manual Revision C 2009381 001 For your notes ...
Страница 99: ...Revsion C DINAMAP ProCare Monitor Service Manual A 1 2009381 001 A Technical Specifications ...
Страница 100: ...A 2 DINAMAP ProCare Monitor Service Manual Revision C 2009381 001 For your notes ...
Страница 108: ...A 10 DINAMAP ProCare Monitor Service Manual Revision C 2009381 001 Technical Specifications Specifications ...
Страница 109: ...Revision C DINAMAP ProCare Monitor Service Manual B 1 2009381 001 B Appropriate Use of NIBP Simulators ...
Страница 110: ...B 2 DINAMAP ProCare Monitor Service Manual Revision C 2009381 001 Appropriate Use of NIBP Simulators For your notes ...
Страница 115: ...Revision C DINAMAP ProCare Monitor Service Manual C 1 2009381 001 C Electromagnetic Compatibility EMC ...
Страница 116: ...C 2 DINAMAP ProCare Monitor Service Manual Revision C 2009381 001 Electromagnetic Compatibility EMC For your notes ...
Страница 133: ......