Revision C
DINAMAP
®
ProCare Monitor Service Manual
4-13
2009381-001
Maintenance:
Parameter Level Functional Testing
6. Close valve on bulb and slowly inflate pressure to 200 mmHg (using the
manometer as reference).
7. Press
Menu
button when pressure reads exactly 200 mmHg to save
calibration setting.
8. Press
Cycle
button until
6
appears in the
min
window.
9. Press and hold
Menu
button until monitor beeps, which acknowledges that
data was saved.
10. Turn the Monitor off.
Overpressure Verification
NOTE:
:
To enter service mode press and hold
Cycle
button while pressing the
On/Off
button.
1. Wait a few seconds after entering service mode. Press
Cycle
button so that
the min window changes from
0
to
1
.
2. Use the inflation bulb to inflate close to 300 mmHg. Slowly inflate (1 to 2
mmHg/sec) until valve opens and pressure is released.
3. Record and verify pressure at which valve opens.
4. Press
Cycle
button and
2
should appear in the
min
window.
5. Use the inflation bulb to inflate close to 150 mmHg. Slowly inflate (1 to 2
mmHg/sec) until valve opens and pressure is released.
6. Record and verify pressure at which valve opens.
7. Turn unit off.
Button Testing
1. Disconnect the cuff/hose assembly and power on the unit.
2. Press
Inflate/Stop
button.
3. Verify a BP determination has been initiated.
4. Block pump port and verify
E80
alarm.
5. Verify flashing red indicator of
Silence
button.
6. Press
Silence
button, verify alarm has been removed.
7. Press
Alarm
button several times, verify unit cycles through all alarm
settings (i.e., SYS, DIA, SpO
2
).
8. Turn unit off.
LED Tests
1. Power on the Monitor.
2. During the power-up self-test verify all 7 segment LED display segments and
all discrete LEDs (except CHARGING LEDs) illuminate.
3. Repeat power up cycle until all LEDs are checked.
Содержание DINAMAP ProCare
Страница 2: ......
Страница 11: ...Revision C DINAMAP ProCare Monitor Service Manual 1 1 2009381 001 1 Introduction ...
Страница 12: ...1 2 DINAMAP ProCare Monitor Service Manual Revision C 2009381 001 For your notes ...
Страница 23: ...Revision C DINAMAP ProCare Monitor Service Manual 2 1 2009381 001 2 Equipment Overview ...
Страница 24: ...2 2 DINAMAP ProCare Monitor Service Manual Revision C 2009381 001 For your notes ...
Страница 38: ...2 16 DINAMAP ProCare Monitor Service Manual Revision C 2009381 001 Equipment Overview Functional Description ...
Страница 39: ...Revision C DINAMAP ProCare Patient Monitor Service Manual 3 1 2009381 001 3 Installation ...
Страница 40: ...3 2 DINAMAP ProCare Patient Monitor Service Manual Revision C 2009381 001 For your notes ...
Страница 53: ...Revision C DINAMAP ProCare Monitor Service Manual 4 1 2009381 001 4 Maintenance ...
Страница 54: ...4 2 DINAMAP ProCare Monitor Service Manual Revision C 2009381 001 For your notes ...
Страница 71: ...Revision C DINAMAP ProCare Monitor Service Manual 5 1 2009381 001 5 Troubleshooting ...
Страница 72: ...5 2 DINAMAP ProCare Monitor Service Manual Revision C 2009381 001 For your notes ...
Страница 76: ...5 6 DINAMAP ProCare Monitor Service Manual Revision C 2009381 001 Troubleshooting Alarm Code Interpretation ...
Страница 77: ...Revision C DINAMAP ProCare Monitor Service Manual 6 1 2009381 001 6 Parts List Drawings and Replacement ...
Страница 78: ...6 2 DINAMAP ProCare Monitor Service Manual Revision C 2009381 001 For your notes ...
Страница 99: ...Revsion C DINAMAP ProCare Monitor Service Manual A 1 2009381 001 A Technical Specifications ...
Страница 100: ...A 2 DINAMAP ProCare Monitor Service Manual Revision C 2009381 001 For your notes ...
Страница 108: ...A 10 DINAMAP ProCare Monitor Service Manual Revision C 2009381 001 Technical Specifications Specifications ...
Страница 109: ...Revision C DINAMAP ProCare Monitor Service Manual B 1 2009381 001 B Appropriate Use of NIBP Simulators ...
Страница 110: ...B 2 DINAMAP ProCare Monitor Service Manual Revision C 2009381 001 Appropriate Use of NIBP Simulators For your notes ...
Страница 115: ...Revision C DINAMAP ProCare Monitor Service Manual C 1 2009381 001 C Electromagnetic Compatibility EMC ...
Страница 116: ...C 2 DINAMAP ProCare Monitor Service Manual Revision C 2009381 001 Electromagnetic Compatibility EMC For your notes ...
Страница 133: ......