Revision C
DINAMAP
®
ProCare Monitor Service Manual
5-5
2009381-001
Troubleshooting:
Alarm Code Interpretation
E13
Battery
Battery too low to operate the unit
No
Charge battery
External DC source failed
Replace battery
E21
SPO2
Failed SPO2 sensor
Replace SPO2 sensor
E23
SPO2
SPO2 Sensor off finger
Yes
Reposition SPO2 sensor
E25
SPO2
SPO2 No signal
Yes
Reposition SPO2 sensor
E61
Temp
Temp probe failure
No
Replace temperature probe
E63
Temp
Probe disconnected or wrong type
Yes
Check for correct probe
E66
Temp
Temperature probe too hot
No
Replace probe
E80
NIBP
Overpressure
Yes
Check for pinched or occluded
internal tubing
E82
NIBP
Excess air in cuff
Yes
Wrong cuff type
Failed valve on pneumatics
assembly
E83
NIBP
NIBP inflation time out
Yes
Leak in cuff or o-ring in hose
Internal leak in tubing or
pneumatic valve
Pump not turning on
E84
NIBP
NIBP total time out
Yes
Pneumatic assembly failure
E85
NIBP
NIBP level time out
Yes
Cuff placement on patient
Minor leak in cuff
Pneumatic assembly failure
E89
NIBP
NIBP no pulse detected
Yes
Cuff placement on patient
*
Acknowledging an alarm by pressing the Silence button, cancels the alarm.
** Call Technical Support for definitions and instructions.
Error Code
or Problem
Source
Definition
Can be
Silenced?
Probable Source
Содержание DINAMAP ProCare
Страница 2: ......
Страница 11: ...Revision C DINAMAP ProCare Monitor Service Manual 1 1 2009381 001 1 Introduction ...
Страница 12: ...1 2 DINAMAP ProCare Monitor Service Manual Revision C 2009381 001 For your notes ...
Страница 23: ...Revision C DINAMAP ProCare Monitor Service Manual 2 1 2009381 001 2 Equipment Overview ...
Страница 24: ...2 2 DINAMAP ProCare Monitor Service Manual Revision C 2009381 001 For your notes ...
Страница 38: ...2 16 DINAMAP ProCare Monitor Service Manual Revision C 2009381 001 Equipment Overview Functional Description ...
Страница 39: ...Revision C DINAMAP ProCare Patient Monitor Service Manual 3 1 2009381 001 3 Installation ...
Страница 40: ...3 2 DINAMAP ProCare Patient Monitor Service Manual Revision C 2009381 001 For your notes ...
Страница 53: ...Revision C DINAMAP ProCare Monitor Service Manual 4 1 2009381 001 4 Maintenance ...
Страница 54: ...4 2 DINAMAP ProCare Monitor Service Manual Revision C 2009381 001 For your notes ...
Страница 71: ...Revision C DINAMAP ProCare Monitor Service Manual 5 1 2009381 001 5 Troubleshooting ...
Страница 72: ...5 2 DINAMAP ProCare Monitor Service Manual Revision C 2009381 001 For your notes ...
Страница 76: ...5 6 DINAMAP ProCare Monitor Service Manual Revision C 2009381 001 Troubleshooting Alarm Code Interpretation ...
Страница 77: ...Revision C DINAMAP ProCare Monitor Service Manual 6 1 2009381 001 6 Parts List Drawings and Replacement ...
Страница 78: ...6 2 DINAMAP ProCare Monitor Service Manual Revision C 2009381 001 For your notes ...
Страница 99: ...Revsion C DINAMAP ProCare Monitor Service Manual A 1 2009381 001 A Technical Specifications ...
Страница 100: ...A 2 DINAMAP ProCare Monitor Service Manual Revision C 2009381 001 For your notes ...
Страница 108: ...A 10 DINAMAP ProCare Monitor Service Manual Revision C 2009381 001 Technical Specifications Specifications ...
Страница 109: ...Revision C DINAMAP ProCare Monitor Service Manual B 1 2009381 001 B Appropriate Use of NIBP Simulators ...
Страница 110: ...B 2 DINAMAP ProCare Monitor Service Manual Revision C 2009381 001 Appropriate Use of NIBP Simulators For your notes ...
Страница 115: ...Revision C DINAMAP ProCare Monitor Service Manual C 1 2009381 001 C Electromagnetic Compatibility EMC ...
Страница 116: ...C 2 DINAMAP ProCare Monitor Service Manual Revision C 2009381 001 Electromagnetic Compatibility EMC For your notes ...
Страница 133: ......