59
Cleaning and care
Safety precautions
Warnings
• Disconnect equipment from power line before cleaning
• If liquid has accidentally entered the system or its parts,
disconnect the power cord from the power supply and have the
equipment serviced by authorized service personnel.
• Regular preventive maintenance should be carried out annually.
• Do not use unspecified cleaners, materials or chemicals as they
may damage device surfaces, labels, or cause equipment
failures.
• Do not pour or spray any liquid directly on cables or leadwires or
permit fluid to seep into connections or openings.
• Never use conductive solutions, solutions that contain wax, or
wax compounds to clean devices, cables or leadwires.
• Do not immerse any part of the device in liquids or allow liquid to
enter the interior.
• The decision to sterilize must be made per your institution’s
requirements with an awareness of the effect on the integrity of
the cable or leadwire.
• Clean the surface of the probe before and after each patient use.
• CABLE/SENSOR AFTER CARE
−
Do not immerse sensors or patient cables in water, solvents
or cleaning solutions.
−
Do not reuse sensors intended for single patient use.
−
Do not sterilize sensors or patient cables by irradiation,
steam, or ethylene oxide.
−
Clean the surface of the probe before and after each patient
use.
−
Allow sensor and cable to dry completely after cleaning.
Moisture and dirt on the connector can affect the
measurement accuracy.
−
If a probe is damaged in any way, discontinue use
immediately.
−
Inaccurate SpO
2
data can result if a sensor is past its useful
life.
−
A damaged sensor or a sensor soaked in liquid may cause
burns during electrosurgery.
• Since calibration gas contains anesthetic agents, always ensure
sufficient ventilation of the room during calibration.
• Inaccurate readings due to:
−
use of unapproved accessories
−
reuse of single-use accessories
−
affect on pressure strong scavenging suction
• The user may only perform maintenance procedures specifically
described in this manual.
• On the gas module indicates that airway gases should be
calibrated every six months in normal use and every two months
in continuous use.
• Do not autoclave any part of the system with steam or sterilize
with ethylene oxide.
Содержание B20
Страница 2: ......
Страница 18: ...14 E Entropy module 1 Module keys 2 Entropy connector 1 2 ...
Страница 76: ...72 For your notes ...
Страница 78: ...74 For your notes ...
Страница 86: ...82 For your notes ...
Страница 104: ...100 ...
Страница 124: ...120 For your notes ...
Страница 159: ......