127
• No printer selected
−
Select a printer, see the “
“ chapter.
• Patient discharged
−
Discharge patient.
• Patient admitted
−
Admit patient.
• P1 over range/P2 over range
−
Check the patient status.
• P1 under range/P2 under range
−
Check the patient status.
−
Check the cables.
−
Rezero the transducer.
−
Change the module.
−
If the problem persists, contact authorized service personnel.
• Printer error
−
Printer is not working properly. Check that the network
printer is operational.
• Printing...
−
Printing on network printer has been started. Please wait
until the printing is finished.
• Printing ready
−
Monitor has completed sending printing data to the printer.
Please wait until the printing is finished.
• Recorder: out of paper
−
Add paper.
• Recorder: cover open
−
Close the recorder cover.
• Recorder: input voltage high, Recorder: input voltage low
−
Contact authorized service personnel.
• Recorder: thermal array overheat
−
Contact authorized service personnel.
• Recorder module removed
−
Reconnect the recorder module if you need a recorder.
• Recorder: system error 29
−
The message rarely appear when the record information too
large, the current recording will stop also. Please press the
Recorder Start/Stop
key again.
• Recorder: system error X
−
Contact authorized service personnel.
•
Replace Battery A
,
Replace Battery B
−
Replace the battery as soon as possible, see “
• Replace D-Fend
−
Gases: Replace the water trap.
• Restart needed
−
The monitor should be restarted.
• Saving to network
−
Saving trend and information to Network
• Sample line blocked
−
Gases: Change the sampling line and water trap.