109
Changing the units
You can use %, kPa or mmHg as the CO
2
measurement units. The
units can be changed in the
CO2 Setup
menu:
1. Press the
Airway Gas
key.
2. Select
CO2 Setup
-
Unit
.
3. Choose the option.
Selecting the source of respiration
You can select the respiration rate’s source
1. Press the
Airway Gas
key.
2. Select
CO2 Setup
-
Resp Rate Source
.
3. Choose the option.
NOTE: If you select
AUTO,
the rate is calculated from the measured
CO
2
.
OFF/ON the CO
2
measurement
1. Press the
Airway Gas
key.
2. Select
CO2 Setup
-
Measurement
.
3. Choose
OFF
or
ON
Setting alarms
Setting CO
2
alarms
1. Press the
Airway Gas
key.
2. Select
CO2 Setup
-
CO2 Alarm
.
3. Select
Adjust Limits
to set up the limits.
4. Select
EtCO2 Alarm, FiCO2 Alarm
to OFF/ON the limits.
5. You can also select
Alarms Setup
to adjust CO
2
or related
alarms.
Setting respiration rate alarms
1. Press the
Airway Gas
key.
2. Select
CO2 Setup
-
Resp Rate Alarm.
3. Select
Adjust Limits
to set up the limits.
4. Select
RR Alarm
,
to OFF/ON the limits.
5. You can also select
Alarms Setup
to adjust RR or related alarms.