71
Selecting the ECG filter
1. Press the
ECG
key.
2. Select
ECG Setup
-
Filter
, and select the appropriate option:
•
STfilt
: Filters high-frequency artifacts but catches slow ST
changes.
•
Monit
: Filters high-frequency artifacts and slow ST changes.
•
Diagn
: Catches high-frequency changes and slow ST
changes.
Changing the HR source
1. Press the
ECG
key.
2. Select
ECG Setup - HR Source
, and select the appropriate option.
•
AUTO
, which selects the first available HR source
• ECG
• Art
• ABP
• UAC
• Pleth
NOTE:
Art
,
ABP
, and
UAC
display when a related label is selected.
Selecting other HR parameters
You can select what is displayed with heart rate:
1. Press the
ECG
key.
2. Select
ECG Setup
-
Display with HR
, and select the appropriate
option.
• PR
• PVC
•
None
.
Displaying the ECG grid
1. Press the
ECG
key
2. Select
ECG Setup -
Select
Grid
, and select the appropriate
option:
•
ON
: To view ECG wavefroms over gridlines
•
OFF
: To view ECG wavefroms without gridlines
Adjusting the beat sound volume
1. Press the
ECG
key.
2. Select
ECG Setup
-
Beat Sound Volume
.
3. Adjust the volume from
0
to
10
. If you select
0
, there is no audible
sound.
Setting the ECG alarms
1. Press the
ECG
key.
2. Select
ECG Alarms
.
• To set up the limits, select
Adjust Limits
.
• To OFF/ON the HR limits. select
HR Alarm
.
• To OFF/ON the PVC limits, select
PVC Alarm
.
• To adjust ST limits, select
ST Alarms
.
3. You can also select
Alarms Setup
to adjust ECG or related
alarms.