44
Pause audio and alarm reset behaviors
* Alarms are listed as follow:
−
Physiology alarms:
• NIBP Dia high
• NIBP Sys high
• NIBP Mean high
• NIBP Dia low
• NIBP Sys low
• NIBP Mean low
−
Technical alarms:
• Leads off
• Entropy cable off
• Entropy sensor off
• Entropy sensor check failed
• No Entropy sensor
• No SpO2 probe
• SpO2 probe off
• Check SpO2 probe
• No SpO2 pulse
• No Px transducer
• NIBP cuff loose
Action
Result
Indicator
Press
audio pause
key once
• Start a 2 minute audio pause period for all
alarms except the critical breakthrough alarms
(for example, V Tach).
See the “Default configuration worksheet” for
details about breakthrough alarm
• Cease all latched alarms (including message
and light).
• Alarm light: Yes
• Alarm message: Yes and has the audio
pause countdown icon
• Alarm audio: No, except breakthrough
alarms
Press
audio pause
key second time
during audio paused period
• Cease the audio pause state.
• Remove alarms listed below*.
• Alarm light: Yes
• Alarm message: Yes
• Alarm audio: Yes
Press
alarm reset
key
• Pause all active audio alarms for 2 minutes.
• Does not silence any new alarms.
• Cease all latched alarms (including message
and light).
• Cease the audio pause state.
• Alarm light: Yes
• Alarm message: Yes, with audio pause
symbol in the message block
• Alarm audio: No