125
• Check SpO2 probe
−
SpO
2
: Check the sensor and connections.
• Check sample gas out
−
Gases: Remove blockage from the sample gas outlet.
• Condition Battery A, Condition Battery B
−
Condition the battery according to the instructions of the
external charger.
• Creating OCRG snapshot
−
In neonatal mode, the value of HR, SpO
2
or Apnea time is out
of limits.
• Default settings returned.
−
Check your power and reset the settings.
• DEMO MODE
−
DEMO mode is active.
• ECG module error
−
Contact authorized service personnel.
• End of 20 min trend data
−
Check the graphical trends and move the cursor.
• Entropy cable off
−
Connect the Entropy cable to the Entropy module.
• Entropy sensor off
−
Check that the sensor is properly attached to the patient.
• Entropy sensor check failed
−
Check sensor placement and attachment.
−
Press each electrode in the sensor.
−
Replace the sensor.
• Frame temperature high
−
Make sure there is sufficient ventilation.
−
If the problem persists, contact authorized service personnel.
−
Turn off the monitor, wait for it cool down.
• Freq PVCs
−
Check the patient status.
• Incompatible SpO2 Probe
−
Replace the sensor.
−
If the problem persists, contact authorized service personnel.
• InvBP's not zeroed
−
Zero the channel indicated or zero both channels.
• Leads off
−
ECG: Reconnect the disconnected trunk cable, electrode or
leadwire. Change the trunk cable, lead set and module.
• License invalid
−
Contact authorized service personnel to set up the monitor’s
right license again.
• Loading from network
−
Loading mode from network.
• Loading failed
−
Load mode from network failed.