Always check the patient first. If any problem or message persists, contact qualified service personnel.
Messages are listed here in alphabetical order.
• Agent Mixture
−
Mixture of halogenated agents is detected. Check the
ventilator and agent vaporizer settings.
• Alarm setup changed from Central
−
Check the alarm limits and the arrhythmia alarm priorities,
see “
” and “
• Alarms silenced from Central
−
If required, turn on the bedside alarms through
Alarms
Setup
-
Audio ON/OFF
-
Activate Alarms
.
• Alarm volume changed
−
The network connection is lost, and the local alarm volume is
increased. Readjust volume if desired.
• Apnea/No Breath
−
Check the patient status.
−
Check the ventilator and breathing circuit.
• Arrhythmia Paused
−
ECG: Check the patient status. Check the electrodes.
• Art disconnect/ABP disconnect/UAC disconnect
−
IBP: Check the patient status.
−
Check the Art/ABP/UAC line.
• Asystole
−
Check the patient status.
−
Check the electrodes.
• Battery A temperature high, Battery B temperature high
−
Replace the battery.
−
If the problem persists, contact authorized service personnel.
• Battery empty, Battery low
−
Replace the battery, see “
”, or connect
the monitor to power outlet.
• Brady
−
Check the patient status.
• Call service: UMBC error
−
UMBC: Contact authorized service personnel.
• Check D-Fend
−
Check that the water trap is properly attached to the module.
• Check OCRG
−
New OCRG snapshot is created, please check.
• Check NIBP
−
Check the patient status.
−
Check the measurement setup.
−
Check the cuff and hoses.