92
To produce automatic NIBP measurements
To automatically measure NIBP at set time intervals, you must first
set the cycle time before setting the automatic measurements.
You also can configure a custom auto mode to meet the need of
your clinical situation.
Setting cycling intervals
1. Press the
NIBP
key.
2. Select
Cycle Time
.
3. Select the interval time from the list with the Trim Knob.
The possible intervals for autocycling are 1, 2, 3, 4, 5, 10, 15, 20,
30, 60, 90 or 120 minutes and custom option.
NOTE: When you select the custom option, the NIBP measurements
will follow custom mode. You need to configure or verify the custom
mode’s set up before you start the automatic measurements.
Setting custom mode
You can set up to 4 separate steps for auto NIBP determinations by
setting the time interval and how many times this interval is
repeated.
1. Press the
NIBP
key.
2. Select
Custom Setup
.
3. Set up time interval and repeat times for every series.
The cycle time can be set to 1 to 120 min or OFF.
The repeat option can be set to 1 to 25 times or OFF or cont.
Starting and stopping NIBP Auto
• From the keyboard
−
Press the
NIBP Auto On/Off
button in the Command
Board.
−
To stop the measurement, press the
NIBP Auto On/Off
button again.
• From NIBP menu
−
Press the
NIBP
key and select
Start Cycling
.
−
To stop the measurement, select the
Stop Cycling
.
To produce a STAT NIBP measurement
NOTE: Stat NIBP measurement is deactivated in neonatal mode.
To measure NIBP consecutively for 5 minutes:
• From the keyboard
−
To stop the measurement, press the
NIBP Start/Cancel
button again.
• From NIBP menu
−
Press the
NIBP
button and select
Start STAT
.
−
To stop the measurement, select the
Stop STAT
.