41
Alarm activation
Physiological alarms have individual activation criteria as shown in
the table. Alarm annunciation does not depend on case activity.
Checking alarm function
1. Set a parameter alarm limit outside of the current measured
patient values. For example, connect the SpO
2
sensor and adjust
the SpO
2
high limit under the measured SpO
2
values.
2. Confirm that the following alarm notification events occur:
• The audible alarm sounds the correct tone.
• The alarm light illuminates.
• The SpO
2
numeric value flashes in the digit field with the
correct color.
• An alarm printout (if enabled).
3. Audio pause the alarms and confirm that the alarms are paused
and that the left side of the alarm indicator light is a correct
color.
4. Return the parameter alarm limit to the original value.
Parameter
Physiological alarm activation criteria
ECG
Active measurement for 30 seconds.
Impedance
respiration
Active measurement for 30 seconds.
Apnea
After 3 breath within 1 minutes.
SpO2
After successful pulse search.
NIBP
Manual, Auto or Stat mode started.
IBP
Active measurement for 30 seconds.
Temperature
As soon as measurement readings are
available.
Gases
After 1st breath detection.
Entropy
Measurement readings within the preset alarm
limits for 30 seconds.