144
Temperature
Airway gases
Measurement units
° Fahrenheit (F)
° Celsius (C)
Measurement range
10 to 45°C (50 to 113°F)
Measurement accuracy
± 0.1°C without temperature sensor
Display resolution
± 0.1°C at 25 to 45 °C with reusable
probes
Probe types supported
Use only GE Healthcare recommend
temperature YSI probes.
Temperature self-check
At start-up and then every 10 minutes
Probe type time response
Reusable skin temperature
probe:
3 s
Reusable adult central
temperature probe:
6 s
Reusable pediatric central
temperature probe:
4 s
Disposable skin
temperature probe:
3 to 6 s
Disposable central
temperature probe, 12F:
5 to 8 s
Disposable central
temperature probe, 9F:
5 to 8 s
Accuracy specifications apply in normal conditions after a 30 min
warm-up period:
Ambient temperature
18 to 28°C, within ±5°C of calibration
Ambient pressure
660 to 1060 hPa, ±67 hPa of
calibration
Ambient humidity
20 to 80%, within ±20% RH of
calibration
Sampling rate
E-miniC module
E-sCO, E-sCAiO, N-CAiO
module
• 150±25 ml/min (sampling line 2 to
3 m, normal conditions)
• 120±20 ml/min
Warm-up time
E-miniC module
E-sCO, E-sCAiO, N-CAiO
module
• 1 minute for operation, 30 minutes
for full specification
• 1 minute for operation with CO
2
,
O
2
and N
2
O,
5 minutes for operation of
anesthetic agents,
20 minutes for full specification
Respiration rate
Measurement range
E-miniC module
E-sCO, E-sCAiO, N-CAiO
module
• 4 to 80 breaths/min
• 4 to 100 breaths/min