Troubleshooting Flow chart 4
ERROR 301
Frac30 Failure
Cable between
Frac30 and ÄKTA
start connected?
Error message
still displayed?
Run Frac30 diagnostics to perform
indexing of homing.
1. Switch off the instrument.
2. Secure the cable between Frac30
and ÄKTA start .
3. Switch on the instrument.
Is the Drive
sleeve worn out?
(Check)
Proceed with your work.
Replace the Drive sleeve.
Error message
still displayed?
Does the Motor
work properly?
Contact GE Service Engineer.
Contact GE Service Engineer.
Proceed with your work.
Run Frac30 diagnostics. Perform indexing
of homing to check if Motor works
properly.
Note:
This Error message is displayed only
when Frac30 is operated.
Yes
No
No
No
No
Yes
Yes
Yes
No
Yes
142
ÄKTA start Maintenance Manual 29-0603-08 AC