5.14.2
Export system report to USB
Instructions
Follow the instructions below to export the system report to a USB memory stick. Use
the system report in further contacts with GE Service Engineers.
Action
Step
Plug the USB memory stick into the USB port located on ÄKTA start.
1
In the System screen, tap Export System Report to USB.
2
Result: The files INSTLOG.TXT and ERRORLOG.TXT are exported to the USB
memory stick.
Remove the USB memory stick from the USB port and connect it to a com-
puter.
3
Check the content of the system report files INSTLOG.TXT and ERRORLOG.TXT,
see section System report file parameters.
4
System report files parameters
The content of the INSTLOG.TXT file contains the following parameters:
•
ÄKTA start serial no.
•
Firmware version
•
FPGA version
•
Pump run
•
Pump tube run
•
UV LED run time
•
Buffer valve count
•
Wash valve count
•
Outlet valve count
•
Sample valve count
The content of the ERRORLOG.TXT file contains Error codes, for example:
•
501: Over Pressure
•
301: Fraction collector failure
•
603: Illegal operation, restart instrument
ÄKTA start Maintenance Manual 29-0603-08 AC
123
5 Service and settings
5.14 System
5.14.2 Export system report to USB