Action
Step
Set the Cell constant value (Set cell const) for the new Conductivity flow
cell by using the up/down arrows.
3
Tap Save to save the new value.
Note:
•
The Cell constant value is provided on the packaging of the Conductivity
flow cell.
•
If a Cell constant value is not available, calibrate the Conductivity flow
cell as explained in
Section 5.10.2 Temperature sensor calibration, on
and
Section 5.10.3 Conductivity flow cell calibration, on page 98
Set the reference temperature
For the system to recalculate the measured conductivity to conductivity at a
certain reference temperature, enter the temperature in the Set ref temp field.
Note:
Make sure that the checkbox Enable temperature compensation in the
Conductivity:Calibration screen is checked.
Note:
Action
Step
In the Conductivity:Configuration screen, set the reference temperature
(Set ref temp) value in the range 4°C to 35°C.
1
Tap Save to save the new reference temperature.
2
96
ÄKTA start Maintenance Manual 29-0603-08 AC
5 Service and settings
5.10 Conductivity Monitor
5.10.1 Set the Cell constant value