Power failure
The result of a power failure depends on which unit is affected.
will result in...
Power failure to...
•
The run is interrupted immediately.
•
The data collected up to the time of the power fail-
ure is available on the USB memory stick or, if the
system is connected to a computer, UNICORN start.
ÄKTA start
•
The computer with UNICORN start installed shuts
down.
•
On the ÄKTA start Instrument Display, all four touch
buttons will be highlighted.
•
The run is interrupted immediately.
•
Data generated up to 10 seconds before the power
failure can be recovered.
Note:
The UNICORN start client may close down during a
temporary overload of the processor. This may appear
as a computer failure. The run continues and you can
restart the UNICORN start client to regain control.
UNICORN start on a
computer
ÄKTA start Maintenance Manual 29-0603-08 AC
29
2 Safety instructions
2.3 Emergency procedures