5.5.2
Calibration
Parameter description
Description
Parameter
The intended flow rate used for Pump calibration.
Flow Rate
The volume of liquid collected for a certain period of time
at the set flow rate. The Collected Volume value and the
expected value of the volume corresponding to the set
flow rate are used internally for calibration.
Collected Volume
Instructions
Follow the instructions below to calibrate the Pump.
Action
Step
•
Immerse the buffer inlet tubing A in demineralized water.
•
Place the outlet tubing from the Wash valve in a pre-weighed collection
tube.
Note:
Before starting the calibration, prime the flow path with water and make sure
that the outlet tubing where the pumped water is collected is filled with
demineralized water. This ensures that the volume of the collected water
corresponds to the pumped volume.
1
In the Pump screen, tap Calibration.
2
Result: The Calibration screen opens.
78
ÄKTA start Maintenance Manual 29-0603-08 AC
5 Service and settings
5.5 Pump
5.5.2 Calibration