Maintenance action
Interval
Clean the Fraction collector.
Note:
Frac30 is not a spill free design. There will be some spillage when
tubes shift position, depending on the flow rate.
When required
Visually inspect the drive sleeve on the Fraction collector. Re-
place if worn out.
Yearly, or when re-
quried
•
Perform System cleaning, Pump wash A and B, Washout
fractionation tubing.
•
Clean the UV flow cell.
•
Clean the Conductivity flow cell.
When required
Calibrate the Pump.
Weekly, or when
required
Calibrate the UV flow cell.
When required
Calibrate the Conductivity flow cell.
When required
Perform Zero offset calibration for the Pressure sensor.
When required
Replace the inlet filters.
When required
Replace the tubing and connectors.
When required
ÄKTA start Maintenance Manual 29-0603-08 AC
57
5 Service and settings
5.1 Preventive maintenance