7.1
Buffer valve
In case of multiple valve failures, make sure that the valves are not mixed.
Replace the valves one by one by removing one valve at a time from the instru-
ment.
Note:
Required tools
Dimension
Tool
T10
Torx driver
Instructions
Follow the instructions below to remove and replace the Buffer valve.
WARNING
Disconnect power. Always switch off power to the instrument be-
fore replacing any component on the instrument or cleaning the
instrument, unless stated otherwise in the user documentation.
Action
Step
Switch off the instrument by pressing the Power Switch to the O position.
1
Remove the inlet and outlet tubing from the valve ports.
2
ÄKTA start Maintenance Manual 29-0603-08 AC