Total Calibration Event Counter (0x5E)
This reads the total number of calibration events that have been stored in the unit.
Clear all events (0x5F)
Writing zero (0) to this register clears all event counters.
Setting Clock
Please see the table on the following page.
Address
(hex)
Parameter
Function
Data Type
Data range
Access
30
Seconds Time Hi
Seconds Time Hi
Numeric value
0 – 65535
Timer Sec
31
Seconds Time
Low
Seconds Time
Numeric value
0 – 65535
Timer sec
32
Real Time Clock
Year, Month
Read/Set year and
month of RTC
2 Numeric
Values
0-99 year, 1 –
12 month
Timer Struct
33
Real Time Clock
Day, Hour
Read/Set day and
hour of RTC
2 Numeric
Values
1 – 31 day, 0
– 23 hour
34
Real Time Clock
Minute, Second
Read/Set minutes
and seconds of
RTC
2 Numeric
Values
0 – 59 minute,
0 – 59 second
Timer Struct
35 PowerCycleFlag
Read
Power
Cycle
Flag.
Numeric Value
1 – Time Not
Reset; 0 –
Time Was
Reset
Flag
36
Event Index
Event index of
Logged Event
Numeric value
0 - 9
Index
37 Warn
Seconds Time Hi
Seconds Time Hi
for warning event
log entries
Numeric value
0 – 65535
Warn
38
Seconds Time
Low
Seconds Time Low
for warning event
log entries
Numeric value
0 – 65535
Warn
39
Structure time Hi
Hi byte – year, low
byte – month for
warning event log
entries
Numeric value
0 – 65535
Warn
3A
Structure time
Mid
Hi byte – day, low
byte – hour warning
event log entries
Numeric value
0 – 65535
Warn
Содержание Observer-i
Страница 2: ...GASSONIC OBSERVER i ii This page intentionally left blank ...
Страница 16: ...GASSONIC OBSERVER i 16 Figure 7 Internal Earth Terminal ...
Страница 24: ...24 4 G GASSONIC C OBSERV VER i ...
Страница 37: ...GASSONIC OBSERVER i 37 ...
Страница 51: ...GASSONIC OBSERVER i 51 Figure 9 Calibration Routine ...