GASSONIC OBSERVER-i
25
4.4.1. Error / Fault Output
Error/fault conditions are indicated in many ways:
•
On the User Interface display
•
Through the 4-20 mA analog output
•
The Error/Fault relay will de-energize
•
The HART digital information will show an error
•
The Modbus digital information will show an error
•
A fault event will be recorded every 30 seconds
Error / Fault
Display
AO
Modbus
Error relay
User action
Gas
overrides
Low supply
ERV-
0 mA*
0 mA
De-energized
Restore proper
operating voltage
Yes
Acoustic
Error
ERAC
1 mA
**
0 mA
De-energized
Check sound path from
piezo to microphone
Yes
Remote
alarm switch
stuck
EAST
0 mA*
0 mA
De-energized
Check switch wiring
Yes
Remote relay
reset switch
stuck
ERST
0 mA*
0 mA
De-energized
Check switch wiring
Yes
Magnetic
sensors
stuck
EMAG
0 mA*
0 mA
De-energized
Remove magnet
Yes
Internal
voltage error
EINV
0 mA*
0 mA
De-energized
Return to factory
No
Critical
memory error
ECRT
0 mA*
0 mA
De-energized
Return to factory
No
User memory
error
EUSR
0 mA*
0 mA
De-energized
Recycle power and
restore user defaults.
No
HART
memory error
EHRT
0 mA*
0 mA
De-energized
Recycle power and
restore HART
information.
No
Event
memory error
EEVT
0 mA*
0 mA
De-energized
Recycle power event
data maybe accurate.
No
Table 2: Error / Fault Indications
* See acoustic self-test sequence in Section 5.4. ** See HART output current in Table 1.
Low Supply
This fault occurs if the supply voltage at the Gassonic Observer-i drops below +12.5 VDC. When
the supply returns to normal, the Gassonic Observer-i will go back to start up.
Action -
Ensure that the supply voltage is at least +14 VDC at the Gassonic Observer-i.
Содержание Observer-i
Страница 2: ...GASSONIC OBSERVER i ii This page intentionally left blank ...
Страница 16: ...GASSONIC OBSERVER i 16 Figure 7 Internal Earth Terminal ...
Страница 24: ...24 4 G GASSONIC C OBSERV VER i ...
Страница 37: ...GASSONIC OBSERVER i 37 ...
Страница 51: ...GASSONIC OBSERVER i 51 Figure 9 Calibration Routine ...