Revision 4
55
12/15/2018
The Component Table lists all of the compounds in the sample and includes chromatographic
parameters that are used to identify the compounds of interest and perform quantitative
analysis (Section 5.4). In addition, the component table includes parameters that describe
the calibration gas so that a variety of various physical properties of the sample can be
calculated. A configuration can contain two component lists.
The right side of the Configure G.C. tab provides access to several tabs which are used to
set actions that should take place during an analysis, scheduling of analyses and enabling
input and /output ports. . To access each, simply click on the appropriate name.
•
Action List -
lists the various actions that should be performed during an analysis such as
opening or closing a valve or indicating the end of an analysis (Section 5.5). A configuration
can contain two action lists.
•
Streams Setup
- used to indicate the component list and action list that should be used for
each stream. It also includes alarm settings and information about the sampling process
(Section 5.6).
•
Streams Sequencer
- used to select the order of runs in a sequence (Section 5.7).
•
Streams Scheduler
- used to indicate when the various streams should be analyzed.
(Section 5.8)
By default, the analyzer will run the Streams defined in the Streams
Sequencer, but the scheduler can be used to interrupt the sequence. An example of this
would be a scheduled auto calibration
.
•
Digital Inputs
– used to provide information about the four digital inputs (Section 5.9).
•
Digital Outputs
- used to provide information about the four digital outputs (Section 5.10).
•
Analog Inputs
– used to provide information about the four analog inputs (Section 5.11).
•
Analog Outputs
- used to provide information about the four analog outputs (Section 5.12)
•
Global Setting
– used to indicate system identification (Section 5.13).
When a configuration has been established for a given analytical procedure, it is probable
that most of the procedures will be edited on a relatively infrequent basis. Typically, the
configuration is edited when the system is validated, when the nature of the sample changes,
when a new column is installed, when an additional compound must be monitored or if the
schedule should be changed.
On a periodic basis, the system should be calibrated using a gas of known composition. This
procedure is described in Section 6 and may require that some parameters on the
Component Table
be edited.
Configurations can be saved and retrieved as desired.
Содержание AccuChrome MA2963
Страница 21: ...Revision 4 8 12 15 2018 Figure 2 7 The Intrinsically Safe Barrier ...
Страница 24: ...Revision 4 11 12 15 2018 Figure 2 10 4 Minute Analysis Flow Diagram ...
Страница 26: ...Revision 4 13 12 15 2018 Figure 3 1 Physical Dimensions Class 1 Div 2 ACCUCHROME Chromatograph ...
Страница 58: ...Revision 4 45 12 15 2018 Figure 4 24 Chart Commands ...
Страница 65: ...Revision 4 52 12 15 2018 A sample report is shown in Figure 4 32 Figure 4 32 Sample Report ...
Страница 80: ...Revision 4 67 12 15 2018 Figure 5 10 Streams Setup ...
Страница 104: ...Revision 4 91 12 15 2018 Figure 8 2 Enron Mode Short Integers ...
Страница 114: ...Revision 4 101 12 15 2018 SS 2F LE MICRON INLINE FILTER HOUSING SS 2F K4 7 7 MICRON INLINE FILTER ELEMENT ...
Страница 124: ...Revision 4 111 12 15 2018 Figure 12 1 PLGC Power Connections Figure 12 2 Relays Wiring Diagram ...
Страница 125: ...Revision 4 112 12 15 2018 Figure 12 3 Analog Outputs Loop Powered Mode Figure 12 4 Analog Outputs Self Powered Mode ...
Страница 126: ...Revision 4 113 12 15 2018 Figure 12 5 Isolated RS 485 Port Connection Figure 12 6 Isolated RS 232 Connections ...
Страница 129: ...Revision 4 116 12 15 2018 Figure 12 11 Oven RTD and TCD Connections Figure 12 12 Arcnet AC485 Twisted Pair Connection ...
Страница 130: ...Revision 4 117 12 15 2018 Figure 12 13 Oven Heater Connection Figure 12 14 GC Valves Connections ...
Страница 131: ...Revision 4 118 12 15 2018 Figure 12 15 Solenoids 5 to 10 Connections Figure 12 16 Solenoids 1 to 4 Connections ...
Страница 132: ...Revision 4 119 12 15 2018 Figure 12 17 Digital Inputs Connections Figure 12 18 Cabinet Heater Connection ...
Страница 139: ...Revision 4 126 12 15 2018 Section 15 Valco 6 and 10 Port Valve Technical Information 15 1 Valve Operation Instructions ...
Страница 140: ...Revision 4 127 12 15 2018 ...
Страница 141: ...Revision 4 128 12 15 2018 15 2 Valve Maintenance Instructions ...
Страница 142: ...Revision 4 129 12 15 2018 ...