Revision 4
12
12/15/2018
Section 3
Installation
3.1 Receiving the System
When the ACCUCHROME Natural Gas Chromatograph arrives, inspect the packaging for
external signs of damage. If there is any obvious physical damage, immediately contact the
shipping agent and Galvanic Applied Sciences to report the damage and request that the
carrier's agent be present when the unit is unpacked. It is recommended that you retain the
shipping container so that it may be used for future shipment of the unit, if necessary.
3.2 Environmental Requirements
3.2.1 Temperature/Humidity
The ACCUCHROME Natural Gas Chromatograph is designed to be operated at ambient
temperatures from -18
o
C to +60
o
C. The humidity should be between 0 and 95%, non-
condensing. For maximum accuracy and reliability it is recommended that the analyzer be
housed in an environmentally controlled shelter or enclosure.
For outdoor installations, the analyzer should be protected from direct sunlight and rain.
Enclosures are available from Galvanic Applied Sciences.
3.2.2 Space Requirements /Weight
The size and weight of the ACCUCHROME Natural Gas Chromatograph is presented in
Table
3-1. The installation site should provide adequate room for opening the cabinet doors for
maintenance and repair procedures. Complete dimensional information is provided in
Figures 3-1 and 3-2.
When the system is installed, leave 6” between the unit and other devices.
Table 3-1: Space/Weight of the ACCUCHROME Chromatograph
Size
Weight
Class 1, Div 2
33” H x 27”W x 11” D
85 lb
Class 1, Div 1
33” H x 27”W x 11” D
120 lb
Содержание AccuChrome MA2963
Страница 21: ...Revision 4 8 12 15 2018 Figure 2 7 The Intrinsically Safe Barrier ...
Страница 24: ...Revision 4 11 12 15 2018 Figure 2 10 4 Minute Analysis Flow Diagram ...
Страница 26: ...Revision 4 13 12 15 2018 Figure 3 1 Physical Dimensions Class 1 Div 2 ACCUCHROME Chromatograph ...
Страница 58: ...Revision 4 45 12 15 2018 Figure 4 24 Chart Commands ...
Страница 65: ...Revision 4 52 12 15 2018 A sample report is shown in Figure 4 32 Figure 4 32 Sample Report ...
Страница 80: ...Revision 4 67 12 15 2018 Figure 5 10 Streams Setup ...
Страница 104: ...Revision 4 91 12 15 2018 Figure 8 2 Enron Mode Short Integers ...
Страница 114: ...Revision 4 101 12 15 2018 SS 2F LE MICRON INLINE FILTER HOUSING SS 2F K4 7 7 MICRON INLINE FILTER ELEMENT ...
Страница 124: ...Revision 4 111 12 15 2018 Figure 12 1 PLGC Power Connections Figure 12 2 Relays Wiring Diagram ...
Страница 125: ...Revision 4 112 12 15 2018 Figure 12 3 Analog Outputs Loop Powered Mode Figure 12 4 Analog Outputs Self Powered Mode ...
Страница 126: ...Revision 4 113 12 15 2018 Figure 12 5 Isolated RS 485 Port Connection Figure 12 6 Isolated RS 232 Connections ...
Страница 129: ...Revision 4 116 12 15 2018 Figure 12 11 Oven RTD and TCD Connections Figure 12 12 Arcnet AC485 Twisted Pair Connection ...
Страница 130: ...Revision 4 117 12 15 2018 Figure 12 13 Oven Heater Connection Figure 12 14 GC Valves Connections ...
Страница 131: ...Revision 4 118 12 15 2018 Figure 12 15 Solenoids 5 to 10 Connections Figure 12 16 Solenoids 1 to 4 Connections ...
Страница 132: ...Revision 4 119 12 15 2018 Figure 12 17 Digital Inputs Connections Figure 12 18 Cabinet Heater Connection ...
Страница 139: ...Revision 4 126 12 15 2018 Section 15 Valco 6 and 10 Port Valve Technical Information 15 1 Valve Operation Instructions ...
Страница 140: ...Revision 4 127 12 15 2018 ...
Страница 141: ...Revision 4 128 12 15 2018 15 2 Valve Maintenance Instructions ...
Страница 142: ...Revision 4 129 12 15 2018 ...