7.4
Determining if the Chromatograph is Functioning in an Acceptable Manner
There are a number of values that can be checked to see if the system is functioning
properly:
a) Compare the results of the analysis of the calibration gas to the certified value of the
calibration gas. Typically the value for each component should agree to within the
tolerances shown in Table 7-2.
Table 7-2: Typical Allowed Deviation
Component Concentration
Allowed Deviation (ASTM 1945)
0 – 0.01 %
+/- 0.01 %
0.01 – 1 %
+/- 0.04 %
1 – 5 %
+/- 0.07 %
5 – 10 %
+/- 0.08 %
10 – 100%
+/- 0.10%
b) Review the un-normalized total reported by the system for the Reference Gas. Typically
the un-normalized total for the calibration gas should be within 98% to 102%. If the un-
normalized total is not within tolerance, it is possible that the peaks are not being
integrated properly due to a shift in the retention time of one (or more) components.
c) Review the calculated heating value of the calibration gas as analyzed by the
chromatograph with respect to the theoretical heating value of the calibration gas. The
calculated heating value should typically be /- 0.25 BTU.
Содержание AccuChrome MA2963
Страница 21: ...Revision 4 8 12 15 2018 Figure 2 7 The Intrinsically Safe Barrier ...
Страница 24: ...Revision 4 11 12 15 2018 Figure 2 10 4 Minute Analysis Flow Diagram ...
Страница 26: ...Revision 4 13 12 15 2018 Figure 3 1 Physical Dimensions Class 1 Div 2 ACCUCHROME Chromatograph ...
Страница 58: ...Revision 4 45 12 15 2018 Figure 4 24 Chart Commands ...
Страница 65: ...Revision 4 52 12 15 2018 A sample report is shown in Figure 4 32 Figure 4 32 Sample Report ...
Страница 80: ...Revision 4 67 12 15 2018 Figure 5 10 Streams Setup ...
Страница 104: ...Revision 4 91 12 15 2018 Figure 8 2 Enron Mode Short Integers ...
Страница 114: ...Revision 4 101 12 15 2018 SS 2F LE MICRON INLINE FILTER HOUSING SS 2F K4 7 7 MICRON INLINE FILTER ELEMENT ...
Страница 124: ...Revision 4 111 12 15 2018 Figure 12 1 PLGC Power Connections Figure 12 2 Relays Wiring Diagram ...
Страница 125: ...Revision 4 112 12 15 2018 Figure 12 3 Analog Outputs Loop Powered Mode Figure 12 4 Analog Outputs Self Powered Mode ...
Страница 126: ...Revision 4 113 12 15 2018 Figure 12 5 Isolated RS 485 Port Connection Figure 12 6 Isolated RS 232 Connections ...
Страница 129: ...Revision 4 116 12 15 2018 Figure 12 11 Oven RTD and TCD Connections Figure 12 12 Arcnet AC485 Twisted Pair Connection ...
Страница 130: ...Revision 4 117 12 15 2018 Figure 12 13 Oven Heater Connection Figure 12 14 GC Valves Connections ...
Страница 131: ...Revision 4 118 12 15 2018 Figure 12 15 Solenoids 5 to 10 Connections Figure 12 16 Solenoids 1 to 4 Connections ...
Страница 132: ...Revision 4 119 12 15 2018 Figure 12 17 Digital Inputs Connections Figure 12 18 Cabinet Heater Connection ...
Страница 139: ...Revision 4 126 12 15 2018 Section 15 Valco 6 and 10 Port Valve Technical Information 15 1 Valve Operation Instructions ...
Страница 140: ...Revision 4 127 12 15 2018 ...
Страница 141: ...Revision 4 128 12 15 2018 15 2 Valve Maintenance Instructions ...
Страница 142: ...Revision 4 129 12 15 2018 ...