Revision 4
63
12/15/2018
Negative Peak -
if the peak for this gas is a negative peak, check the box.
Old Response Factor -
This is the response factor generated from the previous calibration.
Old Response Time - This is the retention time generated from the previous calibration.
Difference in Response Time -
This is the difference in seconds between the previous
calibration and the current calibration.
Difference in Response Factor -
This is the per cent change in the response factor from the
previous calibration to the current calibration.
5.4.3 Fitting Parameters
This feature is not yet implemented.
5.4.4 Physical Properties of the Calibration Gas
The
Physical Properties of the Calibration Gas
fields (Figure 5-6) shows the calculated
physical properties of the calibration gas. These values can be used for comparison when
analyzing the calibration gas as an unknown in the Reference mode.
Figure 5-6: Physical Properties of the Calibration Gas
Содержание AccuChrome MA2963
Страница 21: ...Revision 4 8 12 15 2018 Figure 2 7 The Intrinsically Safe Barrier ...
Страница 24: ...Revision 4 11 12 15 2018 Figure 2 10 4 Minute Analysis Flow Diagram ...
Страница 26: ...Revision 4 13 12 15 2018 Figure 3 1 Physical Dimensions Class 1 Div 2 ACCUCHROME Chromatograph ...
Страница 58: ...Revision 4 45 12 15 2018 Figure 4 24 Chart Commands ...
Страница 65: ...Revision 4 52 12 15 2018 A sample report is shown in Figure 4 32 Figure 4 32 Sample Report ...
Страница 80: ...Revision 4 67 12 15 2018 Figure 5 10 Streams Setup ...
Страница 104: ...Revision 4 91 12 15 2018 Figure 8 2 Enron Mode Short Integers ...
Страница 114: ...Revision 4 101 12 15 2018 SS 2F LE MICRON INLINE FILTER HOUSING SS 2F K4 7 7 MICRON INLINE FILTER ELEMENT ...
Страница 124: ...Revision 4 111 12 15 2018 Figure 12 1 PLGC Power Connections Figure 12 2 Relays Wiring Diagram ...
Страница 125: ...Revision 4 112 12 15 2018 Figure 12 3 Analog Outputs Loop Powered Mode Figure 12 4 Analog Outputs Self Powered Mode ...
Страница 126: ...Revision 4 113 12 15 2018 Figure 12 5 Isolated RS 485 Port Connection Figure 12 6 Isolated RS 232 Connections ...
Страница 129: ...Revision 4 116 12 15 2018 Figure 12 11 Oven RTD and TCD Connections Figure 12 12 Arcnet AC485 Twisted Pair Connection ...
Страница 130: ...Revision 4 117 12 15 2018 Figure 12 13 Oven Heater Connection Figure 12 14 GC Valves Connections ...
Страница 131: ...Revision 4 118 12 15 2018 Figure 12 15 Solenoids 5 to 10 Connections Figure 12 16 Solenoids 1 to 4 Connections ...
Страница 132: ...Revision 4 119 12 15 2018 Figure 12 17 Digital Inputs Connections Figure 12 18 Cabinet Heater Connection ...
Страница 139: ...Revision 4 126 12 15 2018 Section 15 Valco 6 and 10 Port Valve Technical Information 15 1 Valve Operation Instructions ...
Страница 140: ...Revision 4 127 12 15 2018 ...
Страница 141: ...Revision 4 128 12 15 2018 15 2 Valve Maintenance Instructions ...
Страница 142: ...Revision 4 129 12 15 2018 ...