Revision 4
9
12/15/2018
HELIUM
CARRIER
IN
SAMPLE
SAMPLE LOOP
COLUMN
MEASURING
REFERENCE
VENT
10 TO 100 PSIG
CALIBRATION
IN
JUMPER
1
2
3
4
5
6
7
8
9
SAMPLE
VENT
10
Air Supply
TCD
TO ATMOSPHERE
BUFFER COLUMN
60 PSIG
2.4 Typical GC Oven Arrangements
2.4.1
12-Minute Cycle Time Systems
The oven compartment for a typical 12-minute cycle time ACCUCHROME system is
composed of a Valco model DV22 10 port valve and two columns (some applications may
only require one).
•
The valve, which performs sample injection and back flushing, is actuated by the carrier
gas. It is specifically designed for heavy-duty applications and is rated at 1,000,000
injections before requiring service. The injection volume is controlled by a fixed-volume
sample loop.
•
Chromatography columns are employed to separate the natural gas. In two column natural
gas analysis systems, Column 1 separates all components except C6+, trapping C6+ for a
quick back flush out. Column 2 separates the remaining components. The total analysis
time is about 12 minutes, and the analysis is run isothermally at 70
°
C.
•
The temperature of the chromatograph oven is controlled to minimize retention time shifts
due to ambient temperature changes. The gas flow paths for single column and two column
configurations are shown in Figure 2-8 and Figure 2-9.
Figure 2-8: Single Column Flow Diagram
Содержание AccuChrome MA2963
Страница 21: ...Revision 4 8 12 15 2018 Figure 2 7 The Intrinsically Safe Barrier ...
Страница 24: ...Revision 4 11 12 15 2018 Figure 2 10 4 Minute Analysis Flow Diagram ...
Страница 26: ...Revision 4 13 12 15 2018 Figure 3 1 Physical Dimensions Class 1 Div 2 ACCUCHROME Chromatograph ...
Страница 58: ...Revision 4 45 12 15 2018 Figure 4 24 Chart Commands ...
Страница 65: ...Revision 4 52 12 15 2018 A sample report is shown in Figure 4 32 Figure 4 32 Sample Report ...
Страница 80: ...Revision 4 67 12 15 2018 Figure 5 10 Streams Setup ...
Страница 104: ...Revision 4 91 12 15 2018 Figure 8 2 Enron Mode Short Integers ...
Страница 114: ...Revision 4 101 12 15 2018 SS 2F LE MICRON INLINE FILTER HOUSING SS 2F K4 7 7 MICRON INLINE FILTER ELEMENT ...
Страница 124: ...Revision 4 111 12 15 2018 Figure 12 1 PLGC Power Connections Figure 12 2 Relays Wiring Diagram ...
Страница 125: ...Revision 4 112 12 15 2018 Figure 12 3 Analog Outputs Loop Powered Mode Figure 12 4 Analog Outputs Self Powered Mode ...
Страница 126: ...Revision 4 113 12 15 2018 Figure 12 5 Isolated RS 485 Port Connection Figure 12 6 Isolated RS 232 Connections ...
Страница 129: ...Revision 4 116 12 15 2018 Figure 12 11 Oven RTD and TCD Connections Figure 12 12 Arcnet AC485 Twisted Pair Connection ...
Страница 130: ...Revision 4 117 12 15 2018 Figure 12 13 Oven Heater Connection Figure 12 14 GC Valves Connections ...
Страница 131: ...Revision 4 118 12 15 2018 Figure 12 15 Solenoids 5 to 10 Connections Figure 12 16 Solenoids 1 to 4 Connections ...
Страница 132: ...Revision 4 119 12 15 2018 Figure 12 17 Digital Inputs Connections Figure 12 18 Cabinet Heater Connection ...
Страница 139: ...Revision 4 126 12 15 2018 Section 15 Valco 6 and 10 Port Valve Technical Information 15 1 Valve Operation Instructions ...
Страница 140: ...Revision 4 127 12 15 2018 ...
Страница 141: ...Revision 4 128 12 15 2018 15 2 Valve Maintenance Instructions ...
Страница 142: ...Revision 4 129 12 15 2018 ...