Revision 4
84
12/15/2018
7.2 Role of the Calibration Gas and the Reference Gas
The calibration gas and the reference gas are the same gas, but are handled differently in the
application software. In the
Stream Sequencer
screen, the analyst must indicate the nature of
the gas to be analyzed.
•
Calibration Gas
- If the
Calibration Stream
is selected, the certified composition of the
calibration gas (which is entered in the
Component Table
) is used in conjunction with the
peak area to calculate the response factor for each compound.
•
Reference Gas
- If the
Reference Stream
is selected, the composition of the gas is
determined using the existing response factors and reported. When the Reference Stream
is run, the chromatograph will analyze the calibration gas and simply report the values for
each component without making any adjustment to the response factors. The reported
composition of the gas should be compared to that reported by the supplier of the
calibration gas to ensure that the system is operating on an acceptable basis.
•
Sample Gas -
If the
Sample Stream
is selected, the composition of the gas will be
determined using the existing response factors.
7.3 Performing a Calibration
Before a calibration is performed, it is suggested that the reference stream is analyzed. This
allows the user to determine if the chromatograph is operating in an acceptable manner, or a
fault is present. As an example, if the is significant plugging of a valve or a column, it is
probable that one or more compounds will not elute at the proper time, and the reported
concentrations of various components of the sample will be significantly different that that
expected for the gas.
If a reference stream is not analyzed before a calibration stream, it is
possible that the user will generate an invalid calibration via the calibration
process. Running the Reference run allows the user to identify this
problem and correct it while running the Calibration will mask the problem
by simply adjusting the response factor.
In some cases it may be desirable to perform a daily auto calibration to ensure that the
chromatograph is functioning correctly on a day to day basis. A daily calibration can be set
up to perform the calibration automatically (see Section 5.8). In this case the chromatograph
will automatically make adjustments to the response factors within limits set by the user in the
Allowable Response Factor Deviation
entry of the
Configure G.C
. table .
The chromatograph can also be set up to run the reference gas on a daily basis. In this case
the results of the analysis of the calibration gas are reported and the user should compare
the results to the reported concentrations provided by the supplier of the gas and determine if
the GC is operating correctly based on those results.
Содержание AccuChrome MA2963
Страница 21: ...Revision 4 8 12 15 2018 Figure 2 7 The Intrinsically Safe Barrier ...
Страница 24: ...Revision 4 11 12 15 2018 Figure 2 10 4 Minute Analysis Flow Diagram ...
Страница 26: ...Revision 4 13 12 15 2018 Figure 3 1 Physical Dimensions Class 1 Div 2 ACCUCHROME Chromatograph ...
Страница 58: ...Revision 4 45 12 15 2018 Figure 4 24 Chart Commands ...
Страница 65: ...Revision 4 52 12 15 2018 A sample report is shown in Figure 4 32 Figure 4 32 Sample Report ...
Страница 80: ...Revision 4 67 12 15 2018 Figure 5 10 Streams Setup ...
Страница 104: ...Revision 4 91 12 15 2018 Figure 8 2 Enron Mode Short Integers ...
Страница 114: ...Revision 4 101 12 15 2018 SS 2F LE MICRON INLINE FILTER HOUSING SS 2F K4 7 7 MICRON INLINE FILTER ELEMENT ...
Страница 124: ...Revision 4 111 12 15 2018 Figure 12 1 PLGC Power Connections Figure 12 2 Relays Wiring Diagram ...
Страница 125: ...Revision 4 112 12 15 2018 Figure 12 3 Analog Outputs Loop Powered Mode Figure 12 4 Analog Outputs Self Powered Mode ...
Страница 126: ...Revision 4 113 12 15 2018 Figure 12 5 Isolated RS 485 Port Connection Figure 12 6 Isolated RS 232 Connections ...
Страница 129: ...Revision 4 116 12 15 2018 Figure 12 11 Oven RTD and TCD Connections Figure 12 12 Arcnet AC485 Twisted Pair Connection ...
Страница 130: ...Revision 4 117 12 15 2018 Figure 12 13 Oven Heater Connection Figure 12 14 GC Valves Connections ...
Страница 131: ...Revision 4 118 12 15 2018 Figure 12 15 Solenoids 5 to 10 Connections Figure 12 16 Solenoids 1 to 4 Connections ...
Страница 132: ...Revision 4 119 12 15 2018 Figure 12 17 Digital Inputs Connections Figure 12 18 Cabinet Heater Connection ...
Страница 139: ...Revision 4 126 12 15 2018 Section 15 Valco 6 and 10 Port Valve Technical Information 15 1 Valve Operation Instructions ...
Страница 140: ...Revision 4 127 12 15 2018 ...
Страница 141: ...Revision 4 128 12 15 2018 15 2 Valve Maintenance Instructions ...
Страница 142: ...Revision 4 129 12 15 2018 ...