Revision 5
85
August 2021
Zero Purge (sec)
Y
Config – Timers/Alarms
Time interval during a zero calibration cycle
after switching to back purge and before
calculating the new zero baseline
Zero Hold (sec)
Y
Config – Timers/Alarms
Time interval during a zero calibration cycle
after the analyzer has switched back to
sampling mode and prior to the analyzer
concentration outputs configured for Hold
being updated again.
Integration Time (ms)
Y
Spectrum
Amount of time light is collected by
spectrometer before the intensity is read to
produce a spectrum
Fixed Temperature
(Enable)
Y
Config – Calculation
Enable and select a fixed temperature value
for temperature compensation of results
NOTE: Live cell temperature value will be
used for temperature compensation if
this is not enabled.
Fixed Pressure (Enable)
Y
Config – Calculation
Enable and select a fixed pressure value for
pressure compensation of results
Note: Live cell pressure value will be
used for pressure compensation if this is
not enabled.
Measure Analog Input 3
N
N/A
Enable the measurement of a concentration
sensor connected to Analog Input 3 on the
IO Board. If enabled, this measurement will
be displayed on the
Analysis 2
panel of the
local display.
Analog Input 3 is
Hydrogen
N
N/A
Enable the measurement of Hydrogen via a
sensor connected to Analog Input 3 on the
IO board. If this option and the previous
option are enabled, this measurement will be
displayed on the
Analysis 2
panel of the
local display.
Sample Rate
Y
Config – Calculation
Rate at which the displays and output data
are updated when the analyzer is in
sampling mode
Zero Sample Rate
Y
Config – Calculation
Rate at which the displays and output data
are updated when the analyzer is in back
purge mode.
When the page is first opened, it is opened in Read Only mode, which means that the values
for the displayed parameters can only be viewed, not edited. To edit the parameters, first
click on
Change to Update Mode (
at the top right of the screen. The
LOGIN
dialog box
shown in Figure 59 will then appear.
Figure 59: Login Dialog Box
Enter in the correct password, then click
Login.
To change a parameter:
1. Click on the parameter to be edited. The data entry field for that parameter will change
to a white background.
2. Edit the parameter by typing in a new value. Some parameters such as
Fixed
Temperature Enable
are non-numerical and can be edited by using the pull-down
selection menu.
3. Once editing of the selected field is complete, press Enter
on the keyboard.
4. Change any other parameters to be edited using the procedure in steps 1 to 3.
Содержание 943-TGXeNA
Страница 130: ...Revision 5 130 August 2021 Figure 82 Oven Cabinet Door Removed ...
Страница 131: ...Revision 5 131 August 2021 Figure 83 Control Cabinet Door Removed ...
Страница 132: ...Revision 5 132 August 2021 Figure 84 Power Steam Air Signals Connection Details ...
Страница 133: ...Revision 5 133 August 2021 Figure 85 AC Wiring Schematic ...
Страница 134: ...Revision 5 134 August 2021 Figure 86 DC Signals and Wiring Diagram ...
Страница 135: ...Revision 5 135 August 2021 Figure 87 Flow Diagram ...