Revision 5
103
August 2021
Prior to carrying out spectrometer signal optimization, the analyzer MUST be in back
purge and offline mode.
6.4.1
Automated Optimization
1. Place the analyzer in back purge mode. This can be done by selecting the
Back Purge
field on the
Analysis 1
panel and pressing the
↵
key
to place an
X
in the box OR by
pressing the PURGE key on the keypad.
2. Access the
Spectrum
panel, navigate to the
Optimize Integ Time
button and press
ENTER.
Make sure that the peak height after completion of this operation is ≈28000
A/D counts.
3. Return to the
Analysis 1
Panel.
4. Request a Manual Zero by navigating to the
Manual Zero
button and pressing ENTER.
5. After the Manual Zero is completed, place the analyzer back in Sampling mode by
selecting the
Sample
field and then pressing ENTER to put an x in the box OR pressing
PURGE on the analyzer keypad.
6. Place the analyzer back into ONLINE mode by toggling the button on the
Analysis 2
panel.
6.4.2 Manual Optimization
1. Place the analyzer in back purge mode. This can be done by selecting the
Back Purge
field on the
Analysis 1
panel and pressing the ENTER key
to place an
X
in the box
OR by pressing the PURGE key on the keypad.
2. Access the
Spectrum
panel.
3. If the peak level is between 25,000 and 28,000 then no adjustment is required at this
time.
4. If adjustment is necessary, optimize the peak signal height to
≈28
,000 counts. Navigate
to the
Set Integ. TIme
and press ENTER
to display the Integration Period dialog box
(Section 4.6). Enter a new integration period in the
New Value
field and press ENTER.
After several seconds, the new raw signal will be displayed. New values for the
Integration Period can be entered until the peak height is 28 000 /- 1000
counts.
5. Once the peak height has been optimized, record the peak height value and the
resulting integration period.
7. Return to the
Analysis 1
Panel.
8. Request a Manual Zero by navigating to the
Manual Zero
button and pressing ENTER.
Содержание 943-TGXeNA
Страница 130: ...Revision 5 130 August 2021 Figure 82 Oven Cabinet Door Removed ...
Страница 131: ...Revision 5 131 August 2021 Figure 83 Control Cabinet Door Removed ...
Страница 132: ...Revision 5 132 August 2021 Figure 84 Power Steam Air Signals Connection Details ...
Страница 133: ...Revision 5 133 August 2021 Figure 85 AC Wiring Schematic ...
Страница 134: ...Revision 5 134 August 2021 Figure 86 DC Signals and Wiring Diagram ...
Страница 135: ...Revision 5 135 August 2021 Figure 87 Flow Diagram ...