Revision 5
72
August 2021
Figure 46: Analysis Section - Analysis Page
The
Analysis
page is divided into 5 sections, each displaying different information.
5.2.1.1
Value Display
The
Value Display
section displays concentration data, calculated analysis data, and
analyzer parameter as described in Table 17.
Table 17: Value Display Parameters
Parameter
Description
Air Demand
Present air demand, in % excess air based on currently measured H
2
S and SO
2
concentrations, calculated as per equation in Section 1.1.
H2S
Present calculated concentration of H
2
S in percent
SO2
Present calculated concentration of SO
2
in percent
COS
Present calculated concentration of COS in percent
CS2
Present calculated concentration of CS
2
in percent
NOTE: CS
2
is an optional measurement. Only analyzers configured with this
option will be capable of measuring CS
2
Sulfur Vapour
Present calculated concentration of sulfur vapour in percent
Cell Pressure
Present measured pressure inside the sample measurement cell in mmHg
Cell Temperature
Present measured concentration of the measurement cell block assembly in °C
Probe Temperature
Present measured temperature of the sample gas exiting the probe in °C
Cabinet Temperature
The present measured temperature of the interior of the analyzer’s control cabinet
in °C.
Содержание 943-TGXeNA
Страница 130: ...Revision 5 130 August 2021 Figure 82 Oven Cabinet Door Removed ...
Страница 131: ...Revision 5 131 August 2021 Figure 83 Control Cabinet Door Removed ...
Страница 132: ...Revision 5 132 August 2021 Figure 84 Power Steam Air Signals Connection Details ...
Страница 133: ...Revision 5 133 August 2021 Figure 85 AC Wiring Schematic ...
Страница 134: ...Revision 5 134 August 2021 Figure 86 DC Signals and Wiring Diagram ...
Страница 135: ...Revision 5 135 August 2021 Figure 87 Flow Diagram ...