Revision 5
42
August 2021
door is opened, the purge indicator on the outside of the purge control box will be red and
the Loss of Purge relay will be in the fault state.
The relay contacts of the Loss of Purge relay MUST be connected to a plant
alarm system in a continuously attended location, such as in the plant control
room, in order to maintain the hazardous location certification of the analyzer.