Revision 5
101
August 2021
Figure 70: Replacing the Anti-Solarant Solution
1. Disconnect the source optical fibre from the fluid holder bracket on the UV Lamp
Supply (Item #1 in Figure 70). Let the fibre hang freely and ensure that the end does
not contact anything.
It is suggested that a protective plastic cap is placed onto the end of the
fibre. This fibre is removed from the light path to protect it from unfiltered
exposure to the UV Source Lamp).
2. Loosen the light shield retaining screw on the bottom right-side of the fluid holder and
remove the light shield. (#4 in Figure 70)
3. Loosen the fluid holder locking screw (s) on the left-side of the fluid holder using a
7/64" ball driver.(#2 in Figure 70)
4. Lift the fluid holder from the base by grasping the fluid holder filler plug (#3 in Figure
70).
5. Remove the filler plug from the top of the fluid holder.
6. Dump out the old anti-solarant solution.
The
Anti-Solarant Solution
is not hazardous to the environment.