Figure 78: Spectrometer
1. Place the analyzer in Offline mode by toggling the
Online / Offline
control toggle
(Section 4.3.1) on the
Analysis 2
panel to
Offline
(red).
2. Place the analyzer in back purge mode by pressing the PURGE key on the keypad OR
by placing the X in the
Back Purge
check box on the
Analysis 1
panel.
3. Access the purged control cabinet following proper user company and/or regulatory
agency procedure.
The control cabinet door may NOT be opened while the analyzer is
energized UNLESS the area is known to be non-hazardous. Observe all the
warning labels on the analyzer enclosures.
DO NOT look at the end of disconnected fibres connected to the UV lamp or
place them in contact with skin. The intense UV radiation transmitted
through these fibres can cause eye and skin damage. Fibres should be
capped immediately after disconnection to prevent acidental exposure to
UV radiation.
4. Disconnect the USB cable from the front of the spectrometer (#1).
5. Disconnect the air trickle purge line connected to the bottom left corner of the
spectrometer (if present) (#3)
6. Disconnect the detector fibre connected to the top of the spectrometer by unscrewing
the nut and gently pulling it out of the SMA connector. (#2) Place an orange protective
cap over the end of the fibre to prevent damage.
7. Remove the retaining screws that hold the spectrometer to the mounting bracket. (#4).
Hold a hand under the spectrometer while removing it to prevent it from falling when
the last screw is removed.
1
2
3
4
Содержание 943-TGXeNA
Страница 130: ...Revision 5 130 August 2021 Figure 82 Oven Cabinet Door Removed ...
Страница 131: ...Revision 5 131 August 2021 Figure 83 Control Cabinet Door Removed ...
Страница 132: ...Revision 5 132 August 2021 Figure 84 Power Steam Air Signals Connection Details ...
Страница 133: ...Revision 5 133 August 2021 Figure 85 AC Wiring Schematic ...
Страница 134: ...Revision 5 134 August 2021 Figure 86 DC Signals and Wiring Diagram ...
Страница 135: ...Revision 5 135 August 2021 Figure 87 Flow Diagram ...