RESET ON FAIL
•
Options:
o
No
: You must reset the test before a new test can be started. The test result is
held until reset.
o
Yes
: automatically resets to 'Ready' if a Fail occurs. The result is held for the
Reset Delay time before being cancelled. You may start a new test at any time.
RESET DELAY
•
When automatic Reset On Pass or Reset On Fail is used the result lamps and outputs
are held active at the end of a test for the time set here, then they are automatically
cancelled. If you do not want to automatically cancel the lamps and outputs then set
the time to zero.
A.Z. DELAY
•
The differential pressure measurement always takes a zero reference point at the start
of the test stage of a leak test. There are some rare applications that require a little
more time to settle after the internal valves are operated. This setting allows the
default auto-zero timing to be extended for those applications. Set to zero for default
instrument operation.
STATUS LINE
•
The bottom line of the display is used for status indication. You can choose what is
displayed for normal operation.
o
Blank
: the status line is unused.
o
Counters
: shows pass and fail counters.
o
%Counters
: show pass and fail counters as a percentage
o
Barcode
: shows the last scanned barcode
o
Serial No.
: shows the current serial number.