FCO770 Users Guide Issue 1
Page 69
SERIAL No. MASK
•
The scanned serial number may be checked against a mask pattern to validate it. Any
fixed characters in the serial number should be entered in the mask – these must be
matched exactly. The ‘wild card’ characters shown below may be used to match
numeric or alphabetic characters. If the scanned serial number is not validated it is
rejected.
•
The ‘wild card’ characters are:
o
#
0 to 9 or # required
o
&
A-Z or & required (upper or lower case is accepted)
o
?
A character is required but the actual character is not checked.
o
Space
Character position is not checked
START ON SCAN
•
When the expected barcode data has been read the instrument can be commanded to
automatically start a test.
•
Options: No, Yes.
SCAN ON FAIL
•
When a product fails, a custom fail code may be entered into the result data for
communications, printing, or data logging. This may be used to identify a specific
reason for the failure. The instrument will prompt the operator to scan a fail code.
o
The fail code must be a two-digit number which replaces the Product Result.
o
Codes 00 and 10 are reserved for Product Pass and Product Fail respectively,
so you should not use these.
o
If the instrument is reset when waiting for the fail code to be scanned the
default Product Fail code 10 will be used.
•
Options: No, Yes.
Содержание FCO770