Page 94
FCO770 Users Guide Issue 1
ERROR MESSAGES
A number of error messages can be displayed at the top of the FCO770 Status window in the
event of a system or hardware failure.
MESSAGE:
LEAK ZERO FAULT
Indicates that the zero output of the leak transducer is beyond the automatic zero limits. The
transducer assembly requires re-zeroing; if this is beyond adjustment look for damage or
contact your supplier.
MESSAGE:
PR. ZERO FAULT
Indicates that the zero output of the pressure transducer is beyond the allowable 7.5% limit
when attempting zero using the Calibrate Pressure menu. This is often simply an indication
that the air supply had not been turned off when zeroing. If this is not the case then the
transducer assembly requires re-zeroing; if this is beyond adjustment look for damage or
contact your supplier.
MESSAGE:
F.V. STUCK OFF
The Fill Valve may be stuck in the de-energised position. Check that air is available to drive
the valve. If the instrument has been unused for a long period, an air supply of above 7 Bar
may be required to initially activate the internal valves.
MESSAGE:
F.V. STUCK ON
The Fill Valve may be stuck in the energised position. Contact your supplier.
MESSAGE:
T.V. STUCK OFF
The Test Valve may be stuck in the de-energised position. Check that air is available to drive
the valve. If the instrument has been unused for a long period, an air supply of above 7 Bar
may be required to initially activate the internal valves.
MESSAGE:
T.V. STUCK ON
The Test Valve may be stuck in the energised position. Contact your supplier.
MESSAGE:
P.V. STUCK OFF
The Pressure Valve may be stuck in the de-energised position. Check that air is available to
drive the valve. If the instrument has been unused for a long period, an air supply of above 7
Bar may be required to initially activate the internal valves.
MESSAGE:
P.V. STUCK ON
The Pressure Valve may be stuck in the energised position. Contact your supplier.
MESSAGE:
V.V. STUCK ON
The External Vent Valve may be stuck in the energised position, or the optional valve sensor
may not be fitted or is broken. Check the I/O Configuration menu.
MESSAGE:
V.V. STUCK OFF
The External Vent Valve may be stuck in the de-energised position, or the optional valve
sensor may not be fitted or is broken. Check the I/O Configuration menu.