![Furness controls FCO770 Скачать руководство пользователя страница 72](http://html1.mh-extra.com/html/furness-controls/fco770/fco770_manual_2347256072.webp)
Page 72
FCO770 Users Guide Issue 1
BLANK LINES
•
Prints a number of blank lines after custom printed result data. This allows results
from consecutive tests to be separated and can also be used to align labels when using
a label printer. Up to 20 blank lines may be added.
SEND EOF
•
When a PC or PLC is capturing printed data from the FCO770 it may be useful to add
an “End Of File” character at the end of the printed data. The EOF character may be
represented as <Ctrl Z>, 0x1a, chr$(26) etc.
•
Options: No, Yes.
SEND ON
•
Determines the conditions that allow the data to be printed at the end of a test.
•
Options:
o
RESULT
: The data is printed at the end of the test.
o
PASS
: The data is only printed if the test resulted in a pass.
o
FAIL
: The data is only printed if the test resulted in a fail.
SEND AS
•
Selects the fixed format or custom format for the printed data.
•
Options:
o
FIXED
: the data is sent in a single line of ASCII code in a set format
consisting of step number, result, pressure value, and leak value for each step.
CR is always sent, with LF and EOF optional, as configured.
Step Pass/Fail Test Press. T. P Units
Leak
Leak Units Termination
xxx xx
xxxxxxx
xxxxx
xxxxxxx
xxxx x
x
x
01,
P,
0.400,
kPa , -0.1,
PA
CR LF EOF
o
CUSTOM 1
: This allows the printed result to be customised using the
settings below. The print-out consists of a header block followed by up to
sixteen step results, then an optional footer block and optional blank lines.
Each value is preceded by a label, e.g. “DATE 2009-05-26”.
o
CUSTOM 2:
As for Custom 1, but without the labels, e.g. “2009-05-26”. This
is useful for programmable printers that process the printed output before
printing.
o
ZPL II FORM:
This is for use with Zebra
®
printers using the ZPL II
®
language. The printer must be configured to use a predefined form. The
FCO770’s printed output supplies the data to fill in the form. The custom
settings below determine which values are sent to the form.
The following menu entries define the custom format.
Header section:
LINE 1
→
6
•
Up to six lines of user defined text may be added to the printed header (or footer).
•
Options: Enter a line of text (up to 16 characters long).
•
If the first character of any line is a backslash ‘\’ then the next character is used for
additional control of printed output. This is rarely necessary but may help with
unusual printing requirements. This limits the text line to 14 characters.
Содержание FCO770
Страница 2: ......