Page 64
FCO770 Users Guide Issue 1
test must be a leak or dump test with leak units set for flow (e.g. cc/m).
•
“CAL. LEAK”: Enter the value of the external known calibration leak.
•
“DISPLAY”: Enter the value of the displayed leak result previously recorded.
•
“OK”: The calculated test volume is displayed and you are given the option to save
this value. 'OUT OF RANGE' will be displayed if the figures are not usable.
TIME & DATE
•
The system date and time should always be maintained accurately as it may be
included in printed and logged results.
•
DATE: The date is set in the format yyyy-mm-dd.
•
TIME: The time is set in 24 hour format as hh:mm:ss
PRINT DATA
•
This allows the instruments settings to be printed. Note that this menu option will not
appear if the printer port is disabled in the Printer Settings menu.
•
Options: Sub menu:
o
PRODUCT DATA
o
TEST DATA
o
GLOBAL SETTINGS
o
I/O CONFIGURATION
o
PRINTER SETTINGS
o
COUNTERS
o
LOGGED DATA
(not available for the enhanced logger)
o
SYSTEM INFORMATION
•
Press enter to print the data. For Product Data and Test Data you are able to select the
range to print.
COPY PRODUCT DATA.
•
The settings in one product can easily be replicated into other products. This allows
for faster setting up if many products use similar settings.
o
Select the product to copy from and the product to copy to, then choose OK,
or press
BACK
to exit.
o
Note that the first character of the product i/d of the copied data is changed to
‘*’ to indicate that the data is not the original version and should be edited.