Safety
Electromagnetic Compatibility
Compliance to the Electromagnetic Immunity (1)
The DS-8500 System is intended for use in the electromagnetic environment specified below. The customer or the
user of the DS-8500 System should assure that it is used in such an environment.
Guidance and Manufacturer's Declaration - Electromagnetic Immunity
Immunity Test
IEC 60601-1-2
Test Level
Compliance Level
Electromagnetic Environment -
Guidance
Electrostatic Discharge (ESD)
IEC 61000-4-2
±6kV: contact
±8kV: air
±6kV: contact
±8kV: air
Floors should be wood, concrete or
ceramic tile. If floors are covered with
synthetic material, the relative
humidity should be at least 30%.
Electrical fast transient/burst
IEC 61000-4-4
±2kV: power supply lines
±1kV: input/output lines
±2kV: power supply lines
±1kV: input/output lines
Mains power quality should be that of
a typical commercial or hospital
environment.
Surge
IEC 61000-4-5
±1kV: differential mode
±2kV:common mode
±1kV: differential mode
±2kV:common mode
Mains power quality should be that of
a typical commercial or hospital
environment.
Voltage dips, short interruptions
and voltage variations on power
supply input lines.
IEC 61000-4-11
<5% U
T
*
(>95% dip in U
T
)
for 0.5 cycles
<5% U
T
(>95% dip in U
T
)
for 0.5 cycles
Mains power quality should be that of
a typical commercial or hospital
environment.
If it is required to continuously
operate the DS-8500 System during
power failure, it is recommended to
operate on an uninterrupted power
supply.
40% U
T
(60% dip in U
T
)
for 5 cycles
40% U
T
(60% dip in U
T
)
for 5 cycles
70% U
T
(30% dip in U
T
)
for 25 cycles
70% U
T
(30% dip in U
T
)
for 25 cycles
<5% U
T
(>95% dip in U
T
)
for 5 sec.
<5% U
T
(>95% dip in U
T
)
for 5 sec.
Power Frequency (50/60Hz)
Magnetic Field
IEC 61000-4-8
3A/m
3A/m
Power frequency magnetic fields
should be at levels characteristic of a
typical location in a typical
commercial or hospital environment.
*: U
T
is the AC mains voltage prior to application of the test level.
Содержание 8000 Series
Страница 1: ......
Страница 2: ......
Страница 8: ...6 Contents ...
Страница 10: ...Contents ...
Страница 16: ...Contents ...
Страница 46: ...xxx Safety Electromagnetic Compatibility ...
Страница 48: ...Chapter 1 Installation of the Unit Contents ...
Страница 70: ...Chapter 2 Network System Construction Contents ...
Страница 82: ...2 12 Chapter 2 Network System Construction Wireless Network ...
Страница 84: ...Chapter 3 Using the CF card Contents ...
Страница 90: ...3 6 Chapter 3 Using the CF card Formatting the SD Card ...
Страница 92: ...Chapter 4 Connection to the External Devices Contents ...
Страница 124: ...4 32 Chapter 4 Connection to the External Devices Connection with the Laser Printer ...
Страница 126: ...Chapter 5 Initial Settings Contents ...
Страница 158: ...5 32 Chapter 5 Initial Settings User Mode Registration ...
Страница 160: ...Chapter 6 Setup Item Default Value Contents ...
Страница 200: ...Chapter 7 Replacement Parts Contents ...
Страница 204: ...Chapter 8 Cleaning Disinfecting Storing Contents ...
Страница 212: ...8 8 Chapter 8 Cleaning Disinfecting Storing Cleaning the Equipment and Sensors ...
Страница 214: ...Chapter 9 Maintenance Check Contents ...
Страница 228: ...Index 2 Index ...
Страница 229: ......
Страница 230: ......