xxiii
Safety
Precautions about the Ventilator Monitoring
waveform, pay attention to the displayed BP scale.
If the measurement unit of CO
2
concentration is mmHg and [99mmHg] is selected for "CO
2
(mmHg) Upper
Limit of Transmission" ([Initial Settings]>[System]>[Telemeter]), the CO
2
value of 100mmHg or above will
be transmitted as 99mmHg.
RTC and Data Backup
This equipment is equipped with a built-in clock. When the power of this equipment is turned OFF, this clock
is backed up by a lithium primary battery. If incorrect time is displayed when turning ON the power, a low
battery may be the cause. In such case, contact Fukuda Denshi service representative for replacing the battery.
To protect the data during voltage dip, short interruptions and voltage variations on power supply input lines
or during short duration of power turned OFF, this equipment performs 5-minute (approx.) data backup using
the secondary battery. The data may not be protected if the power is turned OFF within 30 minutes from power
ON.
Precautions about the Ventilator Monitoring
The ventilator alarm sound is set to OFF at factory default setting.
The alarm sound can be turned ON on the Tone/Volume setup screen.
If the DS-8500 System does not generate an alarm even though the ventilator is generating an alarm, or if any
other malfunction occurs, immediately check the ventilator, DS-8500 System, cable, and replace the cable if
necessary. If the malfunction persists, stop using the equipment.
The alarm generation on the DS-8500 System is not guaranteed if the alarm other than specified generates at
the ventilator.
(
Maintenance Manual "Ventilator Measurement and Alarm Input" P4-1)
The ventilator operation should be performed by well-trained and authorized personnel.
When connecting this equipment and the ventilator, use only the specified connection cable.
Verify that this equipment and the ventilator are properly connected.
When connecting the cable, verify that the main power of this equipment and the ventilator is OFF.
When Servo-i/s or SV-300 is connected, RESISTANCE (Insp, Exp), COMPLIANCE, P-V loop, F-V loop
display function is not available.
When FLOW-i is connected, P-V loop, F-V loop display function is not available.
Precautions about the SpO
2
Sensor
Danger of Burn Injury Caused by the SpO
2
Sensor
When monitoring SpO
2
, make sure to use only the specified sensor/relay cable. If any other sensor/relay cable
is used, a high temperature rise of the sensor may place the patient in danger of burns.
If there are any questions regarding the sensor/relay cable use for SpO
2
measurements of this equipment, please
contact Fukuda Denshi service representative.
CAUTION
WARNING
CAUTION
DANGER
Содержание 8000 Series
Страница 1: ......
Страница 2: ......
Страница 8: ...6 Contents ...
Страница 10: ...Contents ...
Страница 16: ...Contents ...
Страница 46: ...xxx Safety Electromagnetic Compatibility ...
Страница 48: ...Chapter 1 Installation of the Unit Contents ...
Страница 70: ...Chapter 2 Network System Construction Contents ...
Страница 82: ...2 12 Chapter 2 Network System Construction Wireless Network ...
Страница 84: ...Chapter 3 Using the CF card Contents ...
Страница 90: ...3 6 Chapter 3 Using the CF card Formatting the SD Card ...
Страница 92: ...Chapter 4 Connection to the External Devices Contents ...
Страница 124: ...4 32 Chapter 4 Connection to the External Devices Connection with the Laser Printer ...
Страница 126: ...Chapter 5 Initial Settings Contents ...
Страница 158: ...5 32 Chapter 5 Initial Settings User Mode Registration ...
Страница 160: ...Chapter 6 Setup Item Default Value Contents ...
Страница 200: ...Chapter 7 Replacement Parts Contents ...
Страница 204: ...Chapter 8 Cleaning Disinfecting Storing Contents ...
Страница 212: ...8 8 Chapter 8 Cleaning Disinfecting Storing Cleaning the Equipment and Sensors ...
Страница 214: ...Chapter 9 Maintenance Check Contents ...
Страница 228: ...Index 2 Index ...
Страница 229: ......
Страница 230: ......