xviii
Safety
Precautions for Using the Equipment
When ECG is not measured, Peak Diastolic Pressure (PDP) cannot be calculated.
The undisplayed BP data (SYS/DIA/Mean) will not generate a BP alarm or be displayed in the tabular trend.
Select the appropriate display type according to the monitoring purpose.
Precautions about the CO
2
Monitoring (HCP-800/HCP-810)
Conduct CO
2
calibration for the following case.
If the CO
2
gas calibration is not performed at a specified interval, CO
2
measurement accuracy may be affected
and also subsequent gas calibration may not be possible.
When the accumulated measurement time exceeds 1,200 hours from the first use.
However, if the first calibration was performed before the accumulated measurement time reaches 720 hours,
another calibration is required when the accumulated measurement time exceeds 1,200 hours from the first
calibration.
When 12 months has elapsed or the accumulated measurement time has exceeded 4,000 hours from the
previous calibration.
When EtCO
2
measurement is not stable or accuracy is degraded compared with other measuring device.
When the patient monitor was not used for a while, or when EtCO
2
was not measured for a while.
Perform the calibration 5 minutes after turning ON the power on the HCP-800/HCP-810.
Do not disconnect the sampling tube during calibration. If disconnected, calibration will cease.
Dispose of calibration gas according to the regulation of each medical institution.
Microstream
®
EtCO
2
sampling tubes are designed for single patient use, and are not to be reused. Do not
attempt to clean, disinfect, sterilize or flush any part of the sampling tube as this can cause damage to the
monitor or lead to cross-infection.
Dispose of sampling tubes according to standard operating procedures or local regulations for the disposal of
contaminated medical waste.
Before use, carefully read the Directions for Use for the Microstream
®
EtCO
2
sampling tube.
Only use Microstream
®
EtCO
2
sampling lines to ensure the monitor functions properly.
Precautions about the CO
2
Monitoring (HPD-800/HPD-810 Gas Unit I/F)
The disposable airway adapter should be opened just before use.
Do not reuse the disposable airway adapter. If sterilized, it will become unusable.
The airway adapter should be attached with the thicker side facing to the patient. If attached oppositely, it may
damage the CO
2
sensor or airway adapter.
Dispose of calibration gas according to the regulation of each medical institution.
Precautions about the BIS Monitoring (HBX-800 BISx I/F Unit)
The conductive parts of sensors and connectors should not contact other conductive parts, including earth.
To reduce the hazard of burns in the high-frequency surgical neutral electrode connection, the BIS sensor
should not be located between the surgical site and the electrosurgical unit return electrode.
The BIS sensor must not be located between defibrillator pads when a defibrillator is used on a patient
connected to the patient monitor.
Continuous impedance checking may need to be disabled if the 1 nA, 128 Hz impedance check signal interferes
with other equipment (e.g., evoked potential monitors).
Considerations when using Electro-Convulsive Therapy (ECT) equipment during BIS monitoring: Place ECT
electrodes as far as possible from the BIS sensor to minimize the effect of interference. Certain ECT equipment
may interfere with the proper function of the BIS monitoring system.
Precautions about the Alarm
Alarm messages will be displayed according to the priority. (Level S > Level H > Level M > Level L> Level N)
For the same alarm level, the alarm message for the newer alarm will be displayed.
CAUTION
CAUTION
CAUTION
CAUTION
Содержание 8000 Series
Страница 1: ......
Страница 2: ......
Страница 8: ...6 Contents ...
Страница 10: ...Contents ...
Страница 16: ...Contents ...
Страница 46: ...xxx Safety Electromagnetic Compatibility ...
Страница 48: ...Chapter 1 Installation of the Unit Contents ...
Страница 70: ...Chapter 2 Network System Construction Contents ...
Страница 82: ...2 12 Chapter 2 Network System Construction Wireless Network ...
Страница 84: ...Chapter 3 Using the CF card Contents ...
Страница 90: ...3 6 Chapter 3 Using the CF card Formatting the SD Card ...
Страница 92: ...Chapter 4 Connection to the External Devices Contents ...
Страница 124: ...4 32 Chapter 4 Connection to the External Devices Connection with the Laser Printer ...
Страница 126: ...Chapter 5 Initial Settings Contents ...
Страница 158: ...5 32 Chapter 5 Initial Settings User Mode Registration ...
Страница 160: ...Chapter 6 Setup Item Default Value Contents ...
Страница 200: ...Chapter 7 Replacement Parts Contents ...
Страница 204: ...Chapter 8 Cleaning Disinfecting Storing Contents ...
Страница 212: ...8 8 Chapter 8 Cleaning Disinfecting Storing Cleaning the Equipment and Sensors ...
Страница 214: ...Chapter 9 Maintenance Check Contents ...
Страница 228: ...Index 2 Index ...
Страница 229: ......
Страница 230: ......