5-18
Chapter 5 Initial Settings
User I/F
1 The trend data and tabular trend data will be stored even after the power is turned OFF. To start monitoring a
new patient, it is necessary to perform discharge procedure on the "Admit/Discharge" menu, and clear the data
of previous patient. When the power has been turned OFF for 30 seconds or more, whether or not to display
the discharge confirmation window can be selected.
[OFF]: The discharge confirmation window will not be displayed and monitoring will be immediately started.
[ON]: The discharge confirmation window will be displayed at power ON if the power has been turned OFF for
30 seconds or more.
2 Set the monitoring condition after the patient has been discharged.
[Admit]: Monitoring will continue even after the discharge operation has been performed.
[Monitor Suspend]: Monitoring will be suspended after the discharge operation. The numeric data display will
be cleared, and alarm generation, NIBP periodic measurement, periodic printing will not be performed.
[Standby]: When a patient is discharged, <Monitor will enter into standby mode.> message will be displayed
for 10 seconds, and automatically enters into standby mode. Pressing the [Cancel] key will cancel the process
to enter into standby mode.
Display during monitoring is suspended
3 [OFF]: At power ON and at discharge, NIBP auto mode will continue even after the patient is discharged
regardless of whether the next patient is admitted or not.
[ON]: At power ON and at discharge, NIBP auto mode will resume by starting a manual measurement for the
newly admitted patient.
Until the NIBP auto mode is resumed or the interval is changed, "Standby" will be displayed inside the NIBP
numeric data box.
4 The backup status when the power is turned ON and when the patient is discharged can be set for each item.
[Backup]: The setting will be backed up.
[Initialize]: Initializes the settings. The initialized settings are as follows.
Selection other than Backup
Item
Setup
Power ON/Discharge
Main Mode
Current Mode
Main Mode 1 to 9
The setting will be initialized to the selected mode.
Extended Display 1
Mode
Current Mode
Extended Display 1:
Mode 1 to 3
The setting will be initialized to the selected mode.
Extended Display 2
Mode
Current Mode
Extended Display 2:
Mode 1 to 3
The setting will be initialized to the selected mode.
Patient Classification
Adult, Child, Neonate
The setting will be initialized to the selected patient classification.
Pacemaker
Not Used
"Not Used" will be set for "Pacemaker".
Alarm Settings
Initialize
The setting will be initialized with the currently selected mode.
Display Configuration
Initialize
The setting will be initialized with the currently selected mode.
ECG1, ECG2 Lead
Initialize
The setting will be initialized with the currently selected mode.
Содержание 8000 Series
Страница 1: ......
Страница 2: ......
Страница 8: ...6 Contents ...
Страница 10: ...Contents ...
Страница 16: ...Contents ...
Страница 46: ...xxx Safety Electromagnetic Compatibility ...
Страница 48: ...Chapter 1 Installation of the Unit Contents ...
Страница 70: ...Chapter 2 Network System Construction Contents ...
Страница 82: ...2 12 Chapter 2 Network System Construction Wireless Network ...
Страница 84: ...Chapter 3 Using the CF card Contents ...
Страница 90: ...3 6 Chapter 3 Using the CF card Formatting the SD Card ...
Страница 92: ...Chapter 4 Connection to the External Devices Contents ...
Страница 124: ...4 32 Chapter 4 Connection to the External Devices Connection with the Laser Printer ...
Страница 126: ...Chapter 5 Initial Settings Contents ...
Страница 158: ...5 32 Chapter 5 Initial Settings User Mode Registration ...
Страница 160: ...Chapter 6 Setup Item Default Value Contents ...
Страница 200: ...Chapter 7 Replacement Parts Contents ...
Страница 204: ...Chapter 8 Cleaning Disinfecting Storing Contents ...
Страница 212: ...8 8 Chapter 8 Cleaning Disinfecting Storing Cleaning the Equipment and Sensors ...
Страница 214: ...Chapter 9 Maintenance Check Contents ...
Страница 228: ...Index 2 Index ...
Страница 229: ......
Страница 230: ......