9-6
Chapter 9 Maintenance Check
Daily and Periodic Check
02
BP Measurement
Error
Set the simulator to 120 mmHg for SYS, 80
mmHg for DIA, 90 mmHg for MAP, and perform
the NIBP measurement.
The displayed BP value should be within 120
mmHg ± 10 mmHg for SYS, 80 mmHg ±10
mmHg for DIA, 90 mmHg ±10 mmHg for
MAP.
03
Pulse Rate
Measurement Error
Set the pulse rate to 60 bpm on the simulator,
and perform the NIBP measurement.
The displayed pulse rate should be within 60
bpm ± 3 bpm.
No
Check Items
Check Procedure
Criteria
9. Temperature
01
Temperature display
error
Input the temperature signal of 37°C from the
simulator, and check the displayed temperature
value on the monitor.
The displayed temperature value should be
within 37°C ± 0.2°C.
No
Check Items
Check Procedure
Criteria
10. Cardiac Output (Blood Temperature, Injectate Temperature)
01
Blood Temperature
Measurement
Input the blood temperature signal of 37°C from
the simulator, and check the displayed blood
temperature value on the monitor.
The displayed blood temperature value
should be within 37°C ± 0.3°C.
02
Injectate Temperature
Measurement
Input the injectate temperature signal of 15°C
from the simulator, and check the displayed
injectate temperature value on the monitor.
The displayed injectate temperature value
should be within 15°C ± 0.5°C.
No
Check Items
Check Procedure
Criteria
11. CO
2
Concentration (Optional)
01
CO
2
Concentration
Calibration
(HCP-800/HCP-810)
Perform calibration according to the procedure
explained in "Chapter 9 Maintenance Check"
(Maintenance Manual).
The calibration should complete without
error.
02
CO
2
Concentration
Measurement
Perform measurement with 5% calibration gas,
and check the displayed CO
2
concentration
value on the monitor.
The displayed CO
2
concentration value
should be within 38 mmHg ± 2 mmHg.
No
Check Items
Check Procedure
Criteria
12. BIS (Optional)
01
Sensor Detection
Connect the signal generator and BIS sensor
simulator, and check that the brain wave is
displayed.
Brain wave should be displayed on the
monitor.
No
Check Items
Check Procedure
Criteria
13. Analog Output
01
ECG Output Accuracy Input 10 Hz sine-wave signal of 1mV (2mVp-p)
amplitude from the simulator, connect the
oscilloscope to the output, and check the
displayed waveform.
The displayed waveform amplitude should
be within 1 V ± 0.1 V (2 Vp-p ± 0.2 Vp-p).
02
BP output accuracy
Input BP signal of 0 mmHg, 100 mmHg, 250
mmHg from the simulator, connect the
oscilloscope to the output, and check the
displayed waveform.
The displayed output voltage should be
within 0 V ± 0.1 V at 0 mmHg,
The displayed output voltage should be
within 1 V ± 0.1 V at 100 mmHg,
No
Check Items
Check Procedure
Criteria
14. Electrical Safety
01
Earth Leakage
Current (NC)
Measure the earth leakage current under
normal condition using a leak measurement
safety tester.
Test according to the test method of IEC 60601-
1.
It should be 5 mA or less.
Содержание 8000 Series
Страница 1: ......
Страница 2: ......
Страница 8: ...6 Contents ...
Страница 10: ...Contents ...
Страница 16: ...Contents ...
Страница 46: ...xxx Safety Electromagnetic Compatibility ...
Страница 48: ...Chapter 1 Installation of the Unit Contents ...
Страница 70: ...Chapter 2 Network System Construction Contents ...
Страница 82: ...2 12 Chapter 2 Network System Construction Wireless Network ...
Страница 84: ...Chapter 3 Using the CF card Contents ...
Страница 90: ...3 6 Chapter 3 Using the CF card Formatting the SD Card ...
Страница 92: ...Chapter 4 Connection to the External Devices Contents ...
Страница 124: ...4 32 Chapter 4 Connection to the External Devices Connection with the Laser Printer ...
Страница 126: ...Chapter 5 Initial Settings Contents ...
Страница 158: ...5 32 Chapter 5 Initial Settings User Mode Registration ...
Страница 160: ...Chapter 6 Setup Item Default Value Contents ...
Страница 200: ...Chapter 7 Replacement Parts Contents ...
Страница 204: ...Chapter 8 Cleaning Disinfecting Storing Contents ...
Страница 212: ...8 8 Chapter 8 Cleaning Disinfecting Storing Cleaning the Equipment and Sensors ...
Страница 214: ...Chapter 9 Maintenance Check Contents ...
Страница 228: ...Index 2 Index ...
Страница 229: ......
Страница 230: ......