5-13
Chapter 5 Initial Settings
Measurement Related Setup
The "Other" menu will be displayed.
1 If outputting the data to PC or other external device using the PC communication function of this system, an
error may be generated to the NIBP measurement time depending on the input interval of the external
device.This system outputs the data at completion of NIBP measurement, and if the external device inputs the
data at 60 minutes interval, 60 minutes time lag will occur. By starting the measurement 5 minutes early, this
time lag between the external device can be minimized.
[ON]: When [60min]/[120min] is selected for the measurement interval, the measurement will start 5 minutes
before the set time.
NOTE
This setting will be disabled when [Meas.] is set for "Periodic Measurement Starting Time"
on the NIBP setup menu.
2 The mean blood pressure value of BP and NIBP can be selected to be measured from the waveform or from
calculation.
[Calc.]: Calculates the mean BP from the following calculation. Mean BP = (Systolic BP + Diastolic BP x 2) /3
[Wave]: In case of NIBP, maximum amplitude of oscillograph will be the mean value (MAP). In case of BP, the
following measurement will be performed.
3 Sets the "Arrhythmia Analysis Filter".
[Disp. Waveform]: The filter selected on "Admit/Discharge" menu or "ECG" setup menu will be set.
[Fixed]: The filter will be fixed to 0.5 Hz to 40 Hz.
NOTE
When [Disp. Waveform] is selected, the filter will be set according to the selection on
[Menu > Parameter > ECG]. If [Diag.] is selected, the filter will be 0.5 Hz to 40 Hz which
is the same with [Fixed].
4 Set the "Synchronized Mark/Tone".
When [Auto] is selected for "Synchronized Mark/Tone", the priority of the synchronizing parameter can be set.
[ECG]: The synchronizing priority will be set in the order of ECG>SpO
2
-1>SpO
2
-2>BP. The synchronized tone
will be set to [ON].
[SpO
2
]: The synchronizing priority will be set in the order of SpO
2
-1>SpO
2
-2>ECG>BP. The synchronized tone
will be set to [ON].
Example of BP Measurement
1: Systolic BP
2: HR interval
3: Mean BP (MEAN)
4: Diastolic BP
1
2
3
5
4
6
7
Содержание 8000 Series
Страница 1: ......
Страница 2: ......
Страница 8: ...6 Contents ...
Страница 10: ...Contents ...
Страница 16: ...Contents ...
Страница 46: ...xxx Safety Electromagnetic Compatibility ...
Страница 48: ...Chapter 1 Installation of the Unit Contents ...
Страница 70: ...Chapter 2 Network System Construction Contents ...
Страница 82: ...2 12 Chapter 2 Network System Construction Wireless Network ...
Страница 84: ...Chapter 3 Using the CF card Contents ...
Страница 90: ...3 6 Chapter 3 Using the CF card Formatting the SD Card ...
Страница 92: ...Chapter 4 Connection to the External Devices Contents ...
Страница 124: ...4 32 Chapter 4 Connection to the External Devices Connection with the Laser Printer ...
Страница 126: ...Chapter 5 Initial Settings Contents ...
Страница 158: ...5 32 Chapter 5 Initial Settings User Mode Registration ...
Страница 160: ...Chapter 6 Setup Item Default Value Contents ...
Страница 200: ...Chapter 7 Replacement Parts Contents ...
Страница 204: ...Chapter 8 Cleaning Disinfecting Storing Contents ...
Страница 212: ...8 8 Chapter 8 Cleaning Disinfecting Storing Cleaning the Equipment and Sensors ...
Страница 214: ...Chapter 9 Maintenance Check Contents ...
Страница 228: ...Index 2 Index ...
Страница 229: ......
Страница 230: ......