Safety
Precautions for Using the Equipment
The gas calibration will automatically start when the MGU-800/810 series multigas unit is connected.
After the warm-up completes, gas calibration will be performed every 4 hours during stable operation.
During warm-up, gas calibration interval will become shorter than during normal operation.
During gas calibration, measurement data will not be updated.
Calibration gas is not required during gas calibration.
Dispose of calibration gas according to the regulation of each medical institution.
Make sure the sampling line and flow sensor is securely connected to prevent any leakage.
The gas leakage inside or outside the equipment will cause measurement error. Make sure to check for leakage
before usage.
An environment with alcoholic vapor may adversely affect the measurement readings.
CO
2
, N
2
O or anesthetic agent in the atmosphere around the MGU-800/810 series may adversely affect the
measurement readings.
SPIRO and ventilator cannot be used simultaneously.
During the warming up process, the date of the last measurement accuracy check cannot be updated. Perform
the measurement accuracy check after the warming up process is completed.
If the accuracy check is performed using a low pressure gas, the accuracy of gas measurement will be reduced.
Make sure to perform the accuracy check using the specified calibration gas before its expiration date.
If the error persists, refer to your nearest service representative.
About the MGU-810 Series
The adult flow sensor dead space is 6.9 mL and the flow resistance is 1.8 cmH
2
O at 60 L/min.
The pediatric flow sensor dead space is 0.75 mL and the flow resistance is 0.9 cmH
2
O at 10 L/min.
Adjust ventilation accordingly.
To prevent condensation, the patient breathing circuit, flow sensor and pressure tubing should not be directly
exposed to cooling equipment such as fans or cooling blankets.
Leakage of gas from the patient breathing system may occur if the pressure or gas sampling lines are not
connected to the MGU-810.
The pressure tube and gas sampling lines of the flow sensor should always be routed from the patient circuit
to the MGU-810 such a way as to avoid kinking.
Flow sensors that have suffered damage to sensor head, tubing or tubing connector must not be used.
If liquid has entered the pressure tubes, it can be removed by gently tapping or shaking the flow sensor.
Precautions about the Patient Admit/Discharge
If monitoring of a new patient is started without discharging the previous patient, data of the new patient will
be added to the data of the previous patient which will result in inaccuracy.
The user mode setting (alarm/display configuration) will remain effective even when the power is turned OFF
or when the patient is discharged. Before monitoring, make sure the current user mode is suitable for the
patient's condition.
Resuming monitoring will also resume the alarm in suspension.
Precautions about the CF/SD Card
Use only the specified CF/SD card.
Use only the CF/SD card formatted with this equipment.
Make sure to turn the power OFF and ON again after the setup data is read from the CF/SD card.
The read setup data will become effective after the power is turned OFF and ON again.
Reading the patient data from the CF/SD card will erase all previous patient data stored in the patient monitor.
Precautions about the Maintenance
When cleaning the touch panel, never use strong-acidic cleaning solution.
CAUTION
CAUTION
CAUTION
Содержание 8000 Series
Страница 1: ......
Страница 2: ......
Страница 8: ...6 Contents ...
Страница 10: ...Contents ...
Страница 16: ...Contents ...
Страница 46: ...xxx Safety Electromagnetic Compatibility ...
Страница 48: ...Chapter 1 Installation of the Unit Contents ...
Страница 70: ...Chapter 2 Network System Construction Contents ...
Страница 82: ...2 12 Chapter 2 Network System Construction Wireless Network ...
Страница 84: ...Chapter 3 Using the CF card Contents ...
Страница 90: ...3 6 Chapter 3 Using the CF card Formatting the SD Card ...
Страница 92: ...Chapter 4 Connection to the External Devices Contents ...
Страница 124: ...4 32 Chapter 4 Connection to the External Devices Connection with the Laser Printer ...
Страница 126: ...Chapter 5 Initial Settings Contents ...
Страница 158: ...5 32 Chapter 5 Initial Settings User Mode Registration ...
Страница 160: ...Chapter 6 Setup Item Default Value Contents ...
Страница 200: ...Chapter 7 Replacement Parts Contents ...
Страница 204: ...Chapter 8 Cleaning Disinfecting Storing Contents ...
Страница 212: ...8 8 Chapter 8 Cleaning Disinfecting Storing Cleaning the Equipment and Sensors ...
Страница 214: ...Chapter 9 Maintenance Check Contents ...
Страница 228: ...Index 2 Index ...
Страница 229: ......
Страница 230: ......